Performance and reproducibility assessment across multiple time-domain near-infrared spectroscopy device replicas
Design and Quality for Biomedical Technologies XV 2022
; 11951, 2022.
Article
in English
| Scopus | ID: covidwho-1846313
ABSTRACT
Reproducibility between different replicas of the same device is an important aspect in the biomedical research field. Multiple replicas of a dual-wavelength, single-channel “NIRSBOX” device have been assembled and characterized. In this work, we present their full performance assessment. Characterization is focused on measurement accuracy, reproducibility, and reliability, following well-defined and widely adopted procedures to assess the quality of diffuse-optics instruments. The results of the performance assessment procedures are promising, demonstrating highly reproducible performances over different time-domain near-infrared spectroscopy devices, a feature of paramount importance when it comes to comparing results from different instruments, e.g. in multicenter studies. © 2022 SPIE
Full text:
Available
Collection:
Databases of international organizations
Database:
Scopus
Language:
English
Journal:
Design and Quality for Biomedical Technologies XV 2022
Year:
2022
Document Type:
Article