Your browser doesn't support javascript.
Evaluation of the Thermo ScientificTM SureTectTMListeria monocytogenes PCR Assay in a Broad Range of Foods and Selected Environmental Surfaces: Pre-Collaborative and Collaborative Study, First Action 2021.05.
Bastin, Benjamin; Thompson, Wesley; Benzinger, M Joseph; Crowley, Erin S; Vandoros, Evangelos J; Leonte, Ana-Maria; Thomas, Daniel; Hughes, Annette; Crabtree, David; Evans, Katharine; Sohier, Daniele.
  • Bastin B; Q Laboratories, Research and Development, 1930 Radcliff Drive, Cincinnati, OH 45204, USA.
  • Thompson W; Q Laboratories, Research and Development, 1930 Radcliff Drive, Cincinnati, OH 45204, USA.
  • Benzinger MJ; Q Laboratories, Research and Development, 1930 Radcliff Drive, Cincinnati, OH 45204, USA.
  • Crowley ES; Q Laboratories, Research and Development, 1930 Radcliff Drive, Cincinnati, OH 45204, USA.
  • Vandoros EJ; Thermo Fisher Scientific, Research and Development, Wade Road, Basingstoke, Hampshire RG24 8PW, UK.
  • Leonte AM; Thermo Fisher Scientific, Research and Development, Wade Road, Basingstoke, Hampshire RG24 8PW, UK.
  • Thomas D; Thermo Fisher Scientific, Research and Development, Wade Road, Basingstoke, Hampshire RG24 8PW, UK.
  • Hughes A; Thermo Fisher Scientific, Research and Development, Wade Road, Basingstoke, Hampshire RG24 8PW, UK.
  • Crabtree D; Thermo Fisher Scientific, Research and Development, Wade Road, Basingstoke, Hampshire RG24 8PW, UK.
  • Evans K; Thermo Fisher Scientific, Research and Development, Wade Road, Basingstoke, Hampshire RG24 8PW, UK.
  • Sohier D; Thermo Fisher Scientific, Research and Development, Wade Road, Basingstoke, Hampshire RG24 8PW, UK.
J AOAC Int ; 105(4): 1069-1091, 2022 Jun 29.
Article in English | MEDLINE | ID: covidwho-1931846
ABSTRACT

BACKGROUND:

The Thermo Scientific™ SureTect™ Listeria monocytogenes PCR Assay uses Solaris reagents for performing PCR for the rapid and specific detection of Listeria monocytogenes in a broad range of foods and selected environmental surfaces.

OBJECTIVE:

To demonstrate reproducibility of the SureTect Listeria monocytogenes PCR Assay in a collaborative study using a challenging matrix, full-fat cottage cheese (25 g). To extend the scope of the method.

METHOD:

In the collaborative study, the candidate method was compared to the United States Food and Drug Administration/Bacteriological Analytical Manual (FDA/BAM) Chapter 10 Listeria reference method. The candidate method used two PCR thermocyclers, the Applied Biosystems™ QuantStudio™ 5 Real-Time PCR instrument (QS5) and the Applied Biosystems 7500 Fast Real-Time PCR instrument (7500 Fast). Eighteen participants from 10 laboratories located within the United States and Europe were solicited for the collaborative study, with 12 participants submitting valid data. Three levels of contamination were evaluated for each matrix. Statistical analysis was conducted according to the probability of detection (POD) statistical model. In addition, to extend the scope, six matrix studies were performed comparing the candidate method to the FDA/BAM reference method. One of these matrixes was also compared to the ISO 11290-12017 Microbiology of the Food Chain-Horizontal Method for the Detection and Enumeration of Listeria monocytogenes and of Listeria spp.-Part 1 Detection Method Reference Method.

RESULTS:

In the collaborative study, the difference in laboratory results indicates equivalence between the candidate method and reference method for the matrix evaluated, and the method demonstrated acceptable inter-laboratory reproducibility as determined in the collaborative evaluation. The two PCR instruments used in the study performed equivalently. All presumptive positives were confirmed via the alternative confirmation procedure. In the pre-collaborative studies, the results showed comparable performances between the candidate method and the reference method for all matrixes tested.

CONCLUSIONS:

Based on the data generated, the method demonstrated acceptable inter-laboratory reproducibility data and statistical analysis. HIGHLIGHTS Due to the COVID-19 pandemic, some participants had to be trained remotely. Additionally, 25 g full-fat cottage cheese is known to be a challenging matrix to test. No unusual cross-contamination or false positive/negative data were reported, highlighting the ease of use, reproducibility, and robustness of the method.
Subject(s)

Full text: Available Collection: International databases Database: MEDLINE Main subject: COVID-19 / Listeria / Listeria monocytogenes Type of study: Experimental Studies / Randomized controlled trials Limits: Humans Country/Region as subject: North America Language: English Journal: J AOAC Int Year: 2022 Document Type: Article Affiliation country: Jaoacint

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: COVID-19 / Listeria / Listeria monocytogenes Type of study: Experimental Studies / Randomized controlled trials Limits: Humans Country/Region as subject: North America Language: English Journal: J AOAC Int Year: 2022 Document Type: Article Affiliation country: Jaoacint