Your browser doesn't support javascript.
Comparative analysis of loop-mediated isothermal amplification combined with microfluidic chip technology and q-PCR in the detection of clinical infectious pathogens.
Zhang, Enqi; Ou, Hongling; Jia, Lianling; Zhang, Wang; Wang, Yemei; Wang, Xinru.
  • Zhang E; The Postgraduate Training Base of Jinzhou Medical University (The PLA Rocket Force Characteristic Medical Center), Beijing, China.
  • Ou H; Department of Clinical Laboratory, The PLA Rocket Force Characteristic Medical Center, Beijing, China.
  • Jia L; Department of Clinical Laboratory, The PLA Rocket Force Characteristic Medical Center, Beijing, China.
  • Zhang W; Department of Clinical Laboratory, The PLA Rocket Force Characteristic Medical Center, Beijing, China.
  • Wang Y; The Postgraduate Training Base of Jinzhou Medical University (The PLA Rocket Force Characteristic Medical Center), Beijing, China.
  • Wang X; Department of Clinical Laboratory, The PLA Rocket Force Characteristic Medical Center, Beijing, China.
J Clin Lab Anal ; 36(8): e24565, 2022 Aug.
Article in English | MEDLINE | ID: covidwho-1999871
ABSTRACT

BACKGROUND:

Rapid diagnosis of infectious pathogens at an early stage is crucial to stabilize the patient's condition, reduce medical costs, and shorten hospital stays. Currently, some point-of-care tests have their own shortcomings. Therefore, we built a microfluidic chip based on loop-mediated isothermal amplification to can quickly and sensitively detect infectious pathogens.

METHODS:

We extracted the DNA of S. aureus, MRSA, Shigella and Klebsiella pneumoniae. Then, the DNA samples were diluted by 10-fold and examined by two

methods:

LAMP-microfluidic chip and q-PCR, the sensitivity of whom was also compared. In addition, the specificity of the two was also examined by detecting the target bacteria and other microorganisms using the same methods. Finally, we extracted and tested the DNA of clinically infected humoral samples to determine the coincidence rate between the two methods and the bacterial culture method.

RESULTS:

For S. aureus, MRSA, Shigella, and Klebsiella pneumoniae, the detection limits of the chip were 2.25 × 103 copies/µl, 5.32 × 103 copies/µl, 2.89 × 103 copies/µl, 6.53 × 102 copies/µl, and the detection limits of q-PCR were 2.25 × 102 copies/µl, 5.32 × 101 copies/µl, 2.89 × 102 copies/µl, 6.53 × 101 copies/µl, respectively. In terms of detection specificity, neither method cross-reacted with other strains. For the detection of infectious humoral samples, the total coincidence rate between the q-PCR and bacterial culture method was 85.7%, 95%, 95%, and 95.5%, and the total coincidence rate between the chip and bacterial culture method was 81%, 95%, 90%, and 86.4%, respectively.

CONCLUSION:

LAMP-microfluidic chip provides a simple, sensitive, specific, convenient, and rapid pathogen detection method for clinically infected humoral samples without relying on expensive equipment or technical personnels.
Subject(s)
Keywords

Full text: Available Collection: International databases Database: MEDLINE Main subject: Staphylococcus aureus / Microfluidics Type of study: Diagnostic study / Prognostic study / Randomized controlled trials Limits: Humans Language: English Journal: J Clin Lab Anal Journal subject: Laboratory Techniques and procedures Year: 2022 Document Type: Article Affiliation country: Jcla.24565

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: Staphylococcus aureus / Microfluidics Type of study: Diagnostic study / Prognostic study / Randomized controlled trials Limits: Humans Language: English Journal: J Clin Lab Anal Journal subject: Laboratory Techniques and procedures Year: 2022 Document Type: Article Affiliation country: Jcla.24565