Sudden Occurrence of Pacemaker Capture Failure during Irreversible Electroporation Ablation for Prostate Cancer in Post-COVID-19 Patient: A Case Report.
Chae, Min Suk; Lee, Nuri; Koh, Hyun Jung.
  • Chae MS; Department of Anesthesiology and Pain Medicine, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul 06591, Korea.
  • Lee N; Department of Anesthesiology and Pain Medicine, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul 06591, Korea.
  • Koh HJ; Department of Anesthesiology and Pain Medicine, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul 06591, Korea.
Medicina (Kaunas) ; 58(10)2022 Oct 07.
Article in English | MEDLINE | ID: covidwho-2066254
ABSTRACT
Irreversible electroporation (IRE) ablation is a novel treatment option for localized prostate cancer. Here, we present a case of an abrupt and fatal arrhythmia during the IRE procedure in a prostate cancer patient with an implanted permanent pacemaker. A 78-year-old male patient with a pacemaker due to sick sinus syndrome and syncope was scheduled for IRE prostate ablation surgery under general anesthesia. He had a history of recovering from coronavirus disease 2019 (COVID-19) after having been vaccinated against it and recovered without sequalae. Pacemaker interrogation and reprogramming to asynchronous AOO mode was carried out before surgery, however, sinus pause occurred repeatedly during ablation pulse delivery. After the first sinus pause of 2.25 s there was a decrease in continuous arterial blood pressure (ABP). During the delivery of the second and third pulses, identical sinus pauses were observed due to failure to capture. However, the atrial-paced rhythm recovered instantly, and vital signs became acceptable. Although sinus pause recovered gradually, the duration thereof was increased by the delivery of more IRE pulses, with a subsequent abrupt decrease seen in blood pressure. The pacemaker was urgently reprogrammed to DOO mode, after which there were no further pacing failures and no hemodynamic adverse events. For patients with pacemakers, close cardiac monitoring in addition to the interrogation of the pacemaker during the electromagnetic interference (EMI) procedure is recommended, especially in the case of having a disease that may aggravate cardiac vulnerability, such as COVID-19.
Subject(s)
Keywords

Full text: Available Collection: International databases Database: MEDLINE Main subject: Pacemaker, Artificial / Prostatic Neoplasms / COVID-19 Type of study: Case report / Prognostic study Topics: Long Covid / Vaccines Limits: Aged / Humans / Male Language: English Journal subject: Medicine Year: 2022 Document Type: Article

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: Pacemaker, Artificial / Prostatic Neoplasms / COVID-19 Type of study: Case report / Prognostic study Topics: Long Covid / Vaccines Limits: Aged / Humans / Male Language: English Journal subject: Medicine Year: 2022 Document Type: Article