Your browser doesn't support javascript.
Development of real-time RT-PCR assays to detect measles virus on the Hologic Panther Fusion® System.
Grant, Jenny; Atapattu, Nadika; Dilcher, Meik; Tan, Chor Ee; Elvy, Juliet; Ussher, James E.
  • Grant J; Southern Community Laboratories, Dunedin, New Zealand. Electronic address: jenny.grant@sclabs.co.nz.
  • Atapattu N; Hologic Australia and New Zealand Pty Ltd, Sydney, Australia.
  • Dilcher M; National Measles and Rubella Laboratory, Christchurch, New Zealand.
  • Tan CE; Southern Community Laboratories, Wellington, New Zealand.
  • Elvy J; Southern Community Laboratories, Dunedin, New Zealand.
  • Ussher JE; Southern Community Laboratories, Dunedin, New Zealand; Department of Microbiology and Immunology, University of Otago, Dunedin, New Zealand.
J Clin Virol ; 159: 105355, 2023 Feb.
Article in English | MEDLINE | ID: covidwho-2159230
ABSTRACT

BACKGROUND:

In 2019, Aotearoa New Zealand (NZ) experienced its worst measles outbreak since 1997. Due to declining childhood vaccination rates since the beginning of the SARS-CoV-2 pandemic, NZ is at serious risk of another major measles outbreak. Our laboratory provides diagnostic services to NZ's Southern region. In 2019 the Southern region experienced the greatest number of cases outside of Auckland and Northland, however we did not have a validated measles PCR assay in our laboratory.

OBJECTIVES:

We sought to develop reverse transcription real-time polymerase chain reaction (RT-PCR) assays for measles on the Hologic Panther Fusion® System by utilising its open access function. STUDY

DESIGN:

Previously published real-time RT-PCR assays were modified and optimised to detect wild-type measles virus (LDT-Mea), and the vaccine strain of measles virus (LDT-MeaVacA), on the Hologic Panther Fusion® System. The assays were clinically validated.

RESULTS:

The LDT-Mea assay has a limit of detection (LoD) of 0.1 CCID50, while the LDT-MeaVacA assay is less sensitive with a LoD of 1 CCID50. Using 27 samples, the clinical sensitivity and specificity was 100% for both assays. Other common respiratory viruses were found not to cross-react with either the LDT-Mea or LDT-MeaVacA assays.

CONCLUSION:

We have successfully adapted and validated for diagnostic use on the Hologic Panther Fusion® System previously published assays to detect wild-type and vaccine strains of the measles virus. The implementation of measles testing on this system will greatly improve the turn-around time for measles testing, and better support the measles public health response, for our region.
Subject(s)
Keywords

Full text: Available Collection: International databases Database: MEDLINE Main subject: COVID-19 / Measles Type of study: Diagnostic study / Observational study / Prognostic study / Randomized controlled trials Topics: Vaccines Limits: Humans Language: English Journal: J Clin Virol Journal subject: Virology Year: 2023 Document Type: Article

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: COVID-19 / Measles Type of study: Diagnostic study / Observational study / Prognostic study / Randomized controlled trials Topics: Vaccines Limits: Humans Language: English Journal: J Clin Virol Journal subject: Virology Year: 2023 Document Type: Article