Rapid PCR kit: lateral flow paper strip with Joule heater for SARS-CoV-2 detection.
Kim, Kihyeun; Lee, Bobin; Park, Jun Hyeok; Park, Ji-Ho; Lee, Ki Joong; Kwak, Tae Joon; Son, Taehwang; Shin, Yong-Beom; Im, Hyungsoon; Kim, Min-Gon.
  • Kim K; Advanced Photonics Research Institute (APRI), Gwangju Institute of Science and Technology (GIST), Gwangju 61005, Republic of Korea.
  • Lee B; Department of Chemistry, Gwangju Institute of Science and Technology (GIST), Gwangju 61005, Republic of Korea. mkim@gist.ac.kr.
  • Park JH; Department of Chemistry, Gwangju Institute of Science and Technology (GIST), Gwangju 61005, Republic of Korea. mkim@gist.ac.kr.
  • Park JH; Department of Chemistry, Gwangju Institute of Science and Technology (GIST), Gwangju 61005, Republic of Korea. mkim@gist.ac.kr.
  • Lee KJ; Bionanotechnology Research Center, Korea Research Institute of Bioscience and Biotechnology (KRIBB), Daejeon 34141, Republic of Korea.
  • Kwak TJ; Center for Systems Biology, Massachusetts General Hospital, Boston, Massachusetts 02114, USA.
  • Son T; Center for Systems Biology, Massachusetts General Hospital, Boston, Massachusetts 02114, USA.
  • Shin YB; Bionanotechnology Research Center, Korea Research Institute of Bioscience and Biotechnology (KRIBB), Daejeon 34141, Republic of Korea.
  • Im H; BioNano Health Guard Research Center (H-GUARD), Daejeon 34141, Republic of Korea.
  • Kim MG; Center for Systems Biology, Massachusetts General Hospital, Boston, Massachusetts 02114, USA.
Mater Horiz ; 10(5): 1697-1704, 2023 05 09.
Article in English | MEDLINE | ID: covidwho-2266462
ABSTRACT
Polymerase chain reaction (PCR)-based diagnostic kits for point-of-care (POC) testing are highly desirable to prevent the spread of infectious diseases. Here, we demonstrate a rapid PCR testing kit that involves integrating a lateral flow paper strip with a nichrome-based thin film heater. The use of a paper membrane as a PCR-solution container results in fast thermocycling without a cooler because the membrane can contain the solution with a high specific surface area where Joule heating is applied. After PCR, amplified products are simultaneously detected at the lateral flow paper strip with the naked eye. Severe acute respiratory syndrome ß-coronavirus RNA can be detected within 30 min after PCR solution injection. This work reveals that the paper membrane can act as not only a capillary flow channel but also as a promising platform for fast PCR and detection.
Subject(s)

Full text: Available Collection: International databases Database: MEDLINE Main subject: SARS-CoV-2 / COVID-19 Type of study: Diagnostic study Limits: Humans Language: English Journal: Mater Horiz Year: 2023 Document Type: Article