Gradient method for accurate affinity determinations.
Anal Biochem
; 667: 115085, 2023 04 15.
Article
in English
| MEDLINE | ID: covidwho-2280080
ABSTRACT
The value of the affinity constants (kd, ka, and KD) that are determined by label free interaction analysis methods are strongly affected by the ligand density at the sensor surface [1]. This paper outlines a new SPR-imaging method that applies a ligand density gradient enabling the analyte response to be extrapolated to Rmax = 0 µRIU. The mass transport limited region is used to determine the analyte concentration. Cumbersome optimization procedures for tuning the ligand density is prevented and surface dependent effects as rebinding, strong biphasic behavior etcetera are minimized. The method can be fully automated for e.g. accurate determination of the quality of antibodies from commercial sources.
Keywords
Full text:
Available
Collection:
International databases
Database:
MEDLINE
Main subject:
Biosensing Techniques
/
Surface Plasmon Resonance
Language:
English
Journal:
Anal Biochem
Year:
2023
Document Type:
Article
Similar
MEDLINE
...
LILACS
LIS