Your browser doesn't support javascript.
Clinical evaluation of AusDiagnostics SARS-CoV-2 multiplex tandem PCR assay.
Attwood, Lucy O; Francis, Michelle J; Hamblin, John; Korman, Tony M; Druce, Julian; Graham, Maryza.
  • Attwood LO; Department of Microbiology, Monash Health, Clayton, VIC, Australia; Monash Infectious Diseases, Monash Health, Clayton, VIC, Australia.
  • Francis MJ; Department of Microbiology, Monash Health, Clayton, VIC, Australia. Electronic address: Michelle.Francis@monashhealth.org.
  • Hamblin J; Department of Microbiology, Monash Health, Clayton, VIC, Australia.
  • Korman TM; Department of Microbiology, Monash Health, Clayton, VIC, Australia; Monash Infectious Diseases, Monash Health, Clayton, VIC, Australia; Faculty of Medicine Nursing & Health Sciences, Monash University, Clayton, VIC, Australia.
  • Druce J; Victorian Infectious Diseases Reference Laboratory, Melbourne Health at the Peter Doherty Institute for Infection and Immunity, Melbourne, VIC, Australia.
  • Graham M; Department of Microbiology, Monash Health, Clayton, VIC, Australia; Monash Infectious Diseases, Monash Health, Clayton, VIC, Australia; Faculty of Medicine Nursing & Health Sciences, Monash University, Clayton, VIC, Australia; Microbiological Diagnostic Unit Public Health Laboratory, University
J Clin Virol ; 128: 104448, 2020 07.
Article in English | MEDLINE | ID: covidwho-306323
ABSTRACT

BACKGROUND:

In the context of the pandemic, the rapid emergency use authorisation of diagnostic assays for SARS-CoV-2 has meant there are few peer-reviewed published studies of clinical performance of commercial assays.

AIMS:

To evaluate the clinical performance of AusDiagnostics respiratory multiplex tandem PCR assay including SARS-CoV-2.

METHODS:

We reviewed the results following implementation of AusDiagnostics respiratory multiplex tandem PCR assay including SARS-CoV-2, and compared with an in-house RT-PCR assay at our State Reference Laboratory.

RESULTS:

Initial validation using AusDiagnostics coronavirus multiplex tandem PCR assay including SARS-CoV-2 demonstrated good concordance with the State Reference Laboratory. After implementing the AusDiagnostics respiratory multiplex tandem PCR assay including SARS-CoV-2, we tested 7839 samples. 127 samples in which SARS-CoV-2 was detected using the AusDiagnostics assay were referred for testing at the State Reference Laboratory, with concordant results in 118/127 (92.9%) of samples. After resolution of discrepancies, 125/127 (98.4%) of AusDiagnostics results were determined to be true positive results. Out of 7839 samples tested for SARS-CoV-2 during this period, only 2 tests (0.02%) were indeterminate results.

CONCLUSION:

The AusDiagnostics respiratory MT-PCR assay is a reliable assay for detection of SARS-CoV-2.
Subject(s)
Keywords

Full text: Available Collection: International databases Database: MEDLINE Main subject: Pneumonia, Viral / Coronavirus Infections / Clinical Laboratory Techniques / Pandemics / Multiplex Polymerase Chain Reaction / Betacoronavirus Type of study: Diagnostic study / Experimental Studies / Observational study / Prognostic study Limits: Adult / Female / Humans / Male / Middle aged Language: English Journal: J Clin Virol Journal subject: Virology Year: 2020 Document Type: Article Affiliation country: J.jcv.2020.104448

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: Pneumonia, Viral / Coronavirus Infections / Clinical Laboratory Techniques / Pandemics / Multiplex Polymerase Chain Reaction / Betacoronavirus Type of study: Diagnostic study / Experimental Studies / Observational study / Prognostic study Limits: Adult / Female / Humans / Male / Middle aged Language: English Journal: J Clin Virol Journal subject: Virology Year: 2020 Document Type: Article Affiliation country: J.jcv.2020.104448