Your browser doesn't support javascript.
Brief comparative evaluation of six open one-step RT-qPCR mastermixes for the detection of SARS-CoV-2 RNA using a Taqman probe.
Haddar, Cyrille; Verhoeven, Paul O; Bourlet, Thomas; Pozzetto, Bruno; Pillet, Sylvie.
  • Haddar C; BioSpeedia, Institut Pasteur, 75015, Paris, France. Electronic address: cyrille.haddar@biospeedia.com.
  • Verhoeven PO; Laboratory of Infectious Agents and Hygiene, University Hospital of Saint-Etienne, and GIMAP (Groupe Immunité des Muqueuses et Agents Pathogènes) EA-3064, Medicine Faculty of Saint-Etienne, Campus Santé-Innovations of Saint-Etienne, Member of University of Lyon, France. Electronic address: paul.verh
  • Bourlet T; Laboratory of Infectious Agents and Hygiene, University Hospital of Saint-Etienne, and GIMAP (Groupe Immunité des Muqueuses et Agents Pathogènes) EA-3064, Medicine Faculty of Saint-Etienne, Campus Santé-Innovations of Saint-Etienne, Member of University of Lyon, France. Electronic address: thomas.bo
  • Pozzetto B; Laboratory of Infectious Agents and Hygiene, University Hospital of Saint-Etienne, and GIMAP (Groupe Immunité des Muqueuses et Agents Pathogènes) EA-3064, Medicine Faculty of Saint-Etienne, Campus Santé-Innovations of Saint-Etienne, Member of University of Lyon, France. Electronic address: bruno.poz
  • Pillet S; Laboratory of Infectious Agents and Hygiene, University Hospital of Saint-Etienne, and GIMAP (Groupe Immunité des Muqueuses et Agents Pathogènes) EA-3064, Medicine Faculty of Saint-Etienne, Campus Santé-Innovations of Saint-Etienne, Member of University of Lyon, France. Electronic address: sylvie.pi
J Clin Virol ; 132: 104636, 2020 11.
Article in English | MEDLINE | ID: covidwho-747685
ABSTRACT

BACKGROUND:

Facing the emergence of a new RNA virus, clinical laboratories are often helpless in the case of a shortage of reagents recommended by Reference Centres.

OBJECTIVES:

To compare five open one step RT-qPCR reagents to the SuperScript™ III Platinum™ One-Step qRT-PCR kit (Invitrogen) considered as the reference one in France at the beginning of the pandemic for detection of the Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) in respiratory specimens by using a laboratory-developed assay targeting the viral RNA dependant RNA polymerase (RdRp) gene. STUDY

DESIGN:

A total of 51 NUCLISENS easyMAG extracts from respiratory specimens was tested on ABI 7500 thermocycler with TaqMan Fast Virus 1-Step Master Mix (Applied Biosystems), Luna® Universal Probe One-Step RT-qPCR Kit (New England Biolabs), GoTaq® Probe 1- Step RT-qPCR System (Promega), LightCycler® Multiplex RNA Virus Master (Roche) and One-step PrimeScript RT-PCR kit (Takara). The CT values obtained using the 5 challenged reagents were compared to those obtained using the reference assay.

RESULTS:

The percentages of concordance were all above 95 %. When comparing the CT values of the 48 extracts exhibiting CT values < 35 obtained with the reference reagent, the results were similar between the reagents although the differences of CT values were quite dispersed.

CONCLUSIONS:

All five reagents can be considered as alternative reagents to the reference for detecting SARS-CoV-2 RNA.
Subject(s)
Keywords

Full text: Available Collection: International databases Database: MEDLINE Main subject: RNA, Viral / Real-Time Polymerase Chain Reaction / SARS-CoV-2 / COVID-19 Type of study: Diagnostic study / Experimental Studies / Prognostic study Limits: Humans Language: English Journal: J Clin Virol Journal subject: Virology Year: 2020 Document Type: Article

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: RNA, Viral / Real-Time Polymerase Chain Reaction / SARS-CoV-2 / COVID-19 Type of study: Diagnostic study / Experimental Studies / Prognostic study Limits: Humans Language: English Journal: J Clin Virol Journal subject: Virology Year: 2020 Document Type: Article