Your browser doesn't support javascript.
Differential Ventilation Using Flow Control Valves as a Potential Bridge to Full Ventilatory Support during the COVID-19 Crisis.
Anesthesiology ; 133(4): 892-904, 2020 10 01.
Article in English | MEDLINE | ID: covidwho-772637
ABSTRACT

BACKGROUND:

During the COVID-19 pandemic, ventilator sharing was suggested to increase availability of mechanical ventilation. The safety and feasibility of ventilator sharing is unknown.

METHODS:

A single ventilator in pressure control mode was used with flow control valves to simultaneously ventilate two patients with different lung compliances. The system was first evaluated using high-fidelity human patient simulator mannequins and then tested for 1 h in two pairs of COVID-19 patients with acute respiratory failure. Patients were matched on positive end-expiratory pressure, fractional inspired oxygen tension, and respiratory rate. Tidal volume and peak airway pressure (PMAX) were recorded from each patient using separate independent spirometers and arterial blood gas samples drawn at 0, 30, and 60 min. The authors assessed acid-base status, oxygenation, tidal volume, and PMAX for each patient. Stability was assessed by calculating the coefficient of variation.

RESULTS:

The valves performed as expected in simulation, providing a stable tidal volume of 400 ml each to two mannequins with compliance ratios varying from 2020 to 2090 ml/cm H2O. The system was then tested in two pairs of patients. Pair 1 was a 49-yr-old woman, ideal body weight 46 kg, and a 55-yr-old man, ideal body weight 64 kg, with lung compliance 27 ml/cm H2O versus 35 ml/cm H2O. The coefficient of variation for tidal volume was 0.2 to 1.7%, and for PMAX 0 to 1.1%. Pair 2 was a 32-yr-old man, ideal body weight 62 kg, and a 56-yr-old woman, ideal body weight 46 kg, with lung compliance 12 ml/cm H2O versus 21 ml/cm H2O. The coefficient of variation for tidal volume was 0.4 to 5.6%, and for PMAX 0 to 2.1%.

CONCLUSIONS:

Differential ventilation using a single ventilator is feasible. Flow control valves enable delivery of stable tidal volume and PMAX similar to those provided by individual ventilators.
Subject(s)

Full text: Available Collection: International databases Database: MEDLINE Main subject: Pneumonia, Viral / Respiration, Artificial / Ventilators, Mechanical / Coronavirus Infections Type of study: Experimental Studies Topics: Long Covid Limits: Adult / Female / Humans / Male / Middle aged Language: English Journal: Anesthesiology Year: 2020 Document Type: Article

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: Pneumonia, Viral / Respiration, Artificial / Ventilators, Mechanical / Coronavirus Infections Type of study: Experimental Studies Topics: Long Covid Limits: Adult / Female / Humans / Male / Middle aged Language: English Journal: Anesthesiology Year: 2020 Document Type: Article