A Novel Approach to Hematology Testing at the Point of Care.
Bransky, Avishay; Larsson, Anders; Aardal, Elisabeth; Ben-Yosef, Yaara; Christenson, Robert H.
  • Bransky A; PixCell Medical Technologies Ltd., Yokneam Ilit, Israel.
  • Larsson A; Department of Medical Sciences, Uppsala University, Uppsala, Sweden.
  • Aardal E; Department of Clinical Chemistry, Linköping University, Linköping, Sweden.
  • Ben-Yosef Y; Department of Biomedical and Clinical Sciences, Linköping University, Linköping, Sweden.
  • Christenson RH; PixCell Medical Technologies Ltd., Yokneam Ilit, Israel.
J Appl Lab Med ; 6(2): 532-542, 2021 03 01.
Article in English | MEDLINE | ID: covidwho-960522
ABSTRACT

BACKGROUND:

The need for rapid point-of-care (POC) diagnostics is now becoming more evident due to the increasing need for timely results and improvement in healthcare service. With the recent COVID-19 pandemic outbreak, POC has become critical in managing the spread of disease. Applicable diagnostics should be readily deployable, easy to use, portable, and accurate so that they fit mobile laboratories, pop-up treatment centers, field hospitals, secluded wards within hospitals, or remote regions, and can be operated by staff with minimal training. Complete blood count (CBC), however, has not been available at the POC in a simple-to-use device until recently. The HemoScreen, which was recently cleared by the FDA for POC use, is a miniature, easy-to-use instrument that uses disposable cartridges and may fill this gap. CONTENT The HemoScreen's analysis method, in contrast to standard laboratory analyzers, is based on machine vision (image-based analysis) and artificial intelligence (AI). We discuss the different methods currently used and compare their results to the vision-based one. The HemoScreen is found to correlate well to laser and impedance-based methods while emphasis is given to mean cell volume (MCV), mean cell hemoglobin (MCH), and platelets (PLT) that demonstrate better correlation when the vision-based method is compared to itself due to the essential differences between the underlying technologies.

SUMMARY:

The HemoScreen analyzer demonstrates lab equivalent performance, tested at different clinical settings and sample characteristics, and might outperform standard techniques in the presence of certain interferences. This new approach to hematology testing has great potential to improve quality of care in a variety of settings.
Subject(s)
Keywords

Full text: Available Collection: International databases Database: MEDLINE Main subject: Point-of-Care Testing / COVID-19 Testing / COVID-19 / Hematologic Tests / Mobile Health Units Type of study: Diagnostic study / Observational study / Prognostic study Limits: Humans Language: English Journal: J Appl Lab Med Year: 2021 Document Type: Article Affiliation country: Jalm

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: Point-of-Care Testing / COVID-19 Testing / COVID-19 / Hematologic Tests / Mobile Health Units Type of study: Diagnostic study / Observational study / Prognostic study Limits: Humans Language: English Journal: J Appl Lab Med Year: 2021 Document Type: Article Affiliation country: Jalm