Performance verification of anti-SARS-CoV-2-specific antibody detection by using four chemiluminescence immunoassay systems.
Wan, Yafang; Li, Zhijie; Wang, Kun; Li, Tian; Liao, Pu.
  • Wan Y; Department of Clinical Laboratory, Chongqing General Hospital, Chongqing, China.
  • Li Z; Department of Clinical Laboratory, Chongqing General Hospital, Chongqing, China.
  • Wang K; Department of Clinical Laboratory, Chongqing General Hospital, Chongqing, China.
  • Li T; Department of Clinical Laboratory, Chongqing General Hospital, Chongqing, China.
  • Liao P; Department of Clinical Laboratory, Chongqing General Hospital, Chongqing, China.
Ann Clin Biochem ; 57(6): 429-434, 2020 11.
Article in English | MEDLINE | ID: covidwho-982712
ABSTRACT

OBJECTIVES:

The purpose of the current study was to evaluate the analytical performance of seven kits for detecting IgM/IgG antibodies against coronavirus (SARS-CoV-2) by using four chemiluminescence immunoassay systems.

METHODS:

Fifty patients diagnosed with SARS-CoV-2 infection and 130 controls without coronavirus infection from the General Hospital of Chongqing were enrolled in the current retrospective study. Four chemiluminescence immunoassay systems, including seven IgM/IgG antibody detection kits for SARS-CoV-2 (A_IgM, A_IgG, B_IgM, B_IgG, C_IgM, C_IgG and D_Ab), were employed to detect antibody concentrations. The chi-square test, the receiver operating characteristic (ROC) curve and Youden's index were determined to verify the cut-off value of each detection system.

RESULTS:

The repeatability verification results of the A, B, C and D systems are all qualified. D_Ab performed best (92% sensitivity and 99.23% specificity), and B_IgM performed worse than the other systems. Except for the A_IgM and C_IgG systems, the optimal diagnostic thresholds and cut-off values of the other kits and their recommendations are inconsistent with each other. B_IgM had the worst AUC, and C_IgG had the best diagnostic accuracy. More importantly, the B_IgG system had the highest false-positive rate for testing patients with AIDS, tumours and pregnancies. The A_IgM system test showed the highest false-positive rates among elderly individuals over 90 years old. COVID-2019 IgM/IgG antibody test systems exhibit performance differences.

CONCLUSIONS:

The Innodx Biotech Total Antibody serum diagnosis kit is the most reliable detection system for anti-SARS-CoV-2 antibodies, which can be used together with nucleic acid tests as an alternative method for SARS-CoV-2 detecting.
Subject(s)
Keywords

Full text: Available Collection: International databases Database: MEDLINE Main subject: Immunoassay / Luminescence / Betacoronavirus / Antibodies, Viral / Antibody Specificity Type of study: Diagnostic study / Experimental Studies / Observational study Limits: Female / Humans / Male / Middle aged Language: English Journal: Ann Clin Biochem Year: 2020 Document Type: Article Affiliation country: 0004563220963847

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: Immunoassay / Luminescence / Betacoronavirus / Antibodies, Viral / Antibody Specificity Type of study: Diagnostic study / Experimental Studies / Observational study Limits: Female / Humans / Male / Middle aged Language: English Journal: Ann Clin Biochem Year: 2020 Document Type: Article Affiliation country: 0004563220963847