ABSTRACT
Background:
New commercially available point-of-care (POC) immunodiagnostic tests are appearing, which may yield rapid results for anti-SARS-CoV-2 antibodies. The aim of this study was to evaluate the diagnostic accuracy of rapid antibody detection tests compared to a validated laboratory-based enzyme-linked immunosorbent assay (ELISA) and to investigate infections amongst healthcare workers (HCWs) after unprotected close contact to COVID-19 patients.Methods:
Blood serum and whole blood of 130 participants were tested with NADAL® COVID-19 IgG/IgM Rapid Test and mö-screen 2019-NCOV Corona Virus Test against a validated ELISA test. Infection status was evaluated using real-time polymerase-chain-reaction.Results:
Acute COVID-19 infection was detected in 2.4% of exposed HCWs. Antibody tests showed an overall frequency of IgG and IgM in 5.3%, with 1.6% asymptomatic infections. The NADAL® test showed a sensitivity (IgM/ IgG) of 100% (100%/ 100%), a specificity (IgM/ IgG) of 98.8% (97.6%/ 100 %), a PPV of 76.9% (57.1%/ 100%), an NPV of 100% (100%/ 100%), and a diagnostic accuracy of 98.8% (97.7%/ 100%). The mö-screen test had a sensitivity (IgM/IgG) of 90.9% (80%/ 100%), a specificity (IgM/IgG) of 98.8% (97.6%/ 100%), a PPV of 76.9% (57.1%/ 100%), an NPV of 99.6% (99.2%/ 100%), and a diagnostic accuracy of 98.5% (96.9%/ 100%).Conclusions:
The frequency of COVID-19 infections in HCWs after unprotected close contact is higher than in the general population of a low-prevalence country. Both POC tests were useful for detecting IgG, but did not perform well for IgM, mainly due to false positive results.