Avaliação da segurança do processamento de fresas intramedulares flexíveis para cirurgia ortopédica / Safety assessment of reprocessing of flexible intramedullary bone reamers for orthopedic surgery / Evaluación de la seguridad del procesamiento de fresas intramedulares flexibles para cirugía ortopédica
de Souza, Rafael Queiroz; Bronzatti, Jeane Aparecida Gonzalez; Laranjeira, Paulo Roberto; Mimica, Lycia Mara Jenné; da Silva, Cely Barreto; Cruz, Aurea Silveira; Graziano, Kazuko Uchikawa.
Rev. SOBECC
; 22(1): 17-22, jan.-mar. 2017.
Artículo
en Inglés, Portugués
| LILACS, BDENF - enfermagem (Brasil), SESSP-IALPROD, SES-SP, SESSP-IALACERVO | ID: biblio-833319
Objetivos:
Avaliar a eficácia de um procedimento operacional padrão para limpeza de fresas intramedulares flexíveis, bem como o alcance da esterilidade, e evi-denciar a citotoxicidade da sujidade residual de uma fresa flexível utilizada na prática assistencial.Métodos:
Fresas intramedulares flexíveis foram pesadas antes do processamento, após contaminação desafio e depois da limpeza. Elas foram contaminadas com Soil Test™, suspensão de Geobacillus stearothermophilus, na concentração de 106 UFC/mL, e farinha de osso bovino. Após processamento, as amostras foram incubadas em meio de cultura por 21 dias. A sujidade residual de uma fresa utilizada na prática foi submetida ao teste de citotoxicidade in vitro.Resultados:
As amostras, embora esterilizadas, apontaram acúmulo de sujidade e o processamento foi ineficaz. A sujidade residual apresentou efeito citotóxico.Conclusão:
Recomenda-se que o design flexível das fresas seja descontinuado pela insegurança no processamento.Objectives:
To assess the efficacy of a standard operational procedure to clean flexible intramedullary bone reamers, as well as the sterilization level, and to show the cytotoxicity of the residual dirtiness of a flexible reamer used in care practice.Methods:
Flexible intramedullary bone reamers were weighed before processing, after challenge contamination and after cleaning. They were contaminated with the Soil Test™, Geobacillus stearothermo-philus suspension, in the concentration of 106 cfu/ml, and bovine bone flour. After processing, the samples were inoculated into a culture medium and incubated for 21 days. Residual dirtiness of a flexible intramedullary bone reamer used in practice was submitted to in vitro cytotoxicity test.Results:
Despite being sterilized, the samples indicate to accumulated dirtiness and the processing was inefficient. Residual dirtiness presented a cytotoxic effect.Conclusion:
It is recommended that the flexible design of reamers is discontinued by the lack of safety of reprocessing.Objetivos:
Evaluar la eficacia de un procedimiento operacional estándar para limpieza de fresas intramedulares flexibles, así como el alcance de la esterilidad, y evidenciar la citotoxicidad de la suciedad residual de una fresa flexible utilizada en la práctica asistencial.Métodos:
Fresas intramedulares flexibles fueron pesadas antes del procesamiento, tras contaminación desafío y después de la limpieza. Fueron contaminadas con Soil Test™, suspensión de Geobacillus stearothermophilus, en la concentración de 106 UFC/mL, y harina de hueso bovino. Tras el procesamiento, las muestras fueron incuba-das en medio de cultura por 21 días. La suciedad residual de una fresa utilizada en la práctica fue sometida al test de citotoxicidad in vitro.Resultados:
Las muestras, aunque esterilizadas, apuntaron acumulación de suciedad y el procesamiento fue ineficaz. La suciedad residual presentó efecto citotóxico.Conclusión:
Se recomienda que el design flexible de las fresas sea descontinuado por la inseguridad en el procesamiento.Asunto(s)
Humanos Ortopedia Cirugía General Equipos y Suministros Ensayo de Materiales Esterilización Seguridad de Equipos
Biblioteca responsable:
BR2499
Ubicación: BR91.2; P
powered by iAHx-2.18-89
Biblioteca Virtual en Salud