Comparison of hemodynamic responses to laryngoscopy and intubation with Truview PCDTM, McGrath® and Macintosh laryngoscope in patients undergoing coronary artery bypass grafting: A randomized prospective study.
SPSS version 17 was used, and appropriate tests applied. P < 0.05 was considered significant.
Results:
Heart rate and diastolic arterial pressure increased at 0 and 1 min of intubation in all the three groups (P < 0.05) while mean arterial pressure increased at 0 min in the MG and TV groups and at 1 min in all three groups (P < 0.05). A significant increase in systolic arterial pressure was only observed in TV group at 1 min (P < 0.05). These hemodynamicparameters returned to baseline by 3 min of intubation in all the groups. The intergroup comparisons of all hemodynamicparameters were not significant at any time of observation. Highest intubation difficulty score was observed with MC (2.16 ± 1.86) as compared with MG (0.55 ± 0.88) and TV (0.42 ± 0.83) groups (P = 0.003 and P = 0.001, respectively). However, duration of laryngoscopy and intubation was significantly less in MC (36.68 ± 16.15 s) as compared with MG (75.25 ± 30.94 s) and TV (60.47 ± 27.45 s) groups (P = 0.000 and 0.003, respectively).