Performance Evaluation of a Point of Care SelexOnTM B-Type Natriuretic Peptide Immunoassay

Dong-Wook JEKARL; Seungok LEE; Hyunyu CHOI; Se-Woon CHO; Hae-il PARK.
Laboratory Medicine Online ; : 207-213, 2020.
Artículo | WPRIM | ID: wpr-836923
Background@#This study was conducted to evaluate the analytical performance of the SelexOnTM B-type natriuretic peptide (BNP) assay (Osang Healthcare Inc., Korea), a new rapid lateral flow immunoassay for point of care (POC) testing using whole blood. @*Methods@#The imprecision, linearity, and method comparison of SelexOnTM BNP assay were evaluated. Two commercial BNP assays, the ADVIA Centaur® BNP (Siemens Health Care diagnostics Inc., USA) and the Triage® BNP assays (Alere, USA), were included for method comparison using 100 whole blood samples from patients. The reference interval was verified using 120 residual samples from health examination participants. @*Results@#The SelexOn BNP had total CVs of 20.3%, 13.3%, and 10.3% in BNP concentrations of 89.44 pg/mL, 480.71 pg/mL, and 1,201.84 pg/mL of control materials, respectively. Linearity was observed from 56 pg/mL to 1544 pg/mL. The SelexOn BNP (y) regression equation was y=0.9706x-21.68 with Centaur BNP (x) (r=0.930) and y=0.7600x+0.0506 with Triage BNP (x) (r=0.845), respectively. The predicted mean difference (%) of the SelexOn BNP at the clinical decision levels (100 pg/mL) was up to 25% lower than the two comparative methods. The SelexOn BNP levels were below 50 pg/mL in 114 (95%) of the 120 samples. @*Conclusions@#The SelexOn BNP using EDTA was developed as a POC test for differential diagnosis or treatment monitoring for acute heart failure. However, clinical decision values must be improved to be compatible with other BNP methods.
Biblioteca responsable: WPRO