The separation was performed on a Shim-pack VP-ODS C18 analytical column (4.6 mm×250 mm, 5 μm), and the mobile phase was formic acid-dicyclo-hexylamine-water (11998, V/V) at a flow rate of 1.0 mL·min-1. The detection wavelength was set at 210 nm and the column temperature was maintained at 40℃. The loading volume was 20 μL. The RP-HPLCmethod and the cation exchange resinmethod adopted by Chinese Pharmacopoeia were both used to determine the content of sodium lactate in compound sodium lactate and glucoseinjection, and the results were compared.
RESULTS:
The linear range of the RP-HPLCmethod was 0.51-6.07 mg·mL-1 (r=0.9999). The average recovery was 99.47% (RSD=0.75%, n=9).