Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Rheumatol Int ; 44(4): 611-620, 2024 Apr.
Article in English | MEDLINE | ID: mdl-38340159

ABSTRACT

Peripheral oxygen saturation (SpO2) using the fingers may have important limitations due to Raynaud's phenomenon and sclerodactyly in patients with systemic sclerosis (SSc). Sensors located at more central body positions may be more accurate as these as less prone to Raynaud attacks. To determine the validity and reliability of the SpO2 measured at the finger, forehead, and earlobe during the 6-Minute Walk Test (6MWT). Eighty two patients with SSc had an arterial line placed while performing the 6MWT. Peripheral oxygen saturation was simultaneously measured by finger, forehead, and earlobe sensors and compared to the arterial oxygen saturation (SaO2) measured before and after the 6MWT. 40 patients repeated the 6MWT one week later to determine re-test reliability. We used Bland-Altman plots to display the agreement between SpO2 and SaO2. The intraclass correlation coefficient for repeated measurement of minimum SpO2 was calculated. The mean difference between SpO2 and SaO2 after the 6MWT was - 3% (SD: ± 5), 0% (SD: ± 2), and 1% (SD: ± 2) for the finger, forehead, and earlobe, respectively. The minimum SpO2 measured at the finger demonstrated the poorest re-test reliability: The ICC (95% CI) showed good agreement using the ear and forehead probe (ICCear = 0.89 [95% CI 0.80; 0.94]; ICCforehead = 0.77 [95% CI 0.60; 0.87]), while a modest reliability was found using the finger probe (ICCfinger = 0.65 95% CI [0.43; 0.80]). SpO2 should be measured using either the earlobe or forehead during the 6MWT in patients with SSc. Clinical Trials.Gov (NCT04650659).


Subject(s)
Oxygen Saturation , Scleroderma, Systemic , Humans , Oxygen , Walk Test , Reproducibility of Results , Scleroderma, Systemic/diagnosis
SELECTION OF CITATIONS
SEARCH DETAIL
...