MAGMA ; 33(2): 257-272, 2020 Apr.
Article in English | MEDLINE | ID: mdl-31487004

ABSTRACT

OBJECTIVE: To provide a review and quantitative analysis of the available fetal MR imaging phantoms. MATERIALS AND METHODS: A literature search was conducted across Pubmed, Google Scholar, and Ryerson University Library databases to identify fetal MR imaging phantoms. Phantoms were graded on a semi-quantitative scale in regards to four evaluation categories: (1) anatomical accuracy in size and shape, (2) dielectric conductivity similar to the simulated tissue, (3) relaxation times similar to simulated tissue, and (4) physiological motion similar to fetal gross body, cardiovascular, and breathing motion. This was followed by statistical analysis to identify significant findings. RESULTS: Seventeen fetal phantoms were identified and had an average overall percentage accuracy of 26%, with anatomical accuracy being satisfied the most (56%) and physiological motion the least (7%). Phantoms constructed using 3D printing were significantly more accurate than conventionally constructed phantoms. DISCUSSION: Currently available fetal phantoms lack accuracy and motion simulation. 3D printing may lead to higher accuracy compared with traditional manufacturing. Future research needs to focus on properly simulating both fetal anatomy and physiological motion to produce a phantom that is appropriate for fetal MRI sequence development and optimization.


Subject(s)
Magnetic Resonance Imaging/methods , Phantoms, Imaging , Prenatal Diagnosis/methods , Computer Simulation , Equipment Design , Female , Heart , Humans , Imaging, Three-Dimensional , Lung , Motion , Printing, Three-Dimensional , Reproducibility of Results , Respiration , Uterus
SELECTION OF CITATIONS
SEARCH DETAIL
...