Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Catheter Cardiovasc Interv ; 98(1): 117-127, 2021 07 01.
Article in English | MEDLINE | ID: mdl-33942962

ABSTRACT

OBJECTIVES: This study sought to obtain in vivo data on a new stent and delivery system specifically designed for implantation in infants with the ability to be enlarged to adult dimensions. BACKGROUND: There are no endovascular stents designed for or approved for use in infants, nor is there a stent capable of being implanted at infant vessel diameters and achieving adult size while maintaining structural integrity. The Minima stent was designed to address these needs. METHODS: This study was performed in 6 piglets who underwent implantation of 22 Minima stents into the following locations: aorta (n = 11), branch pulmonary arteries (n = 6), and central veins (n = 5). RESULTS: Successful deployment occurred in 21/22 attempts. Two instances of post-deployment migration occurred. Stents were re-expanded at 1, 2, 3 and 5 months after implant. All stents regardless of location could be re-dilated to the intended diameter to keep pace with somatic growth (implant diameter 6.9 +/- 1.2 mm; final diameter 16.1 mm +/- 1.4 mm). Histopathology at 1 and 5 months demonstrated widely patent vessel lumens with stent apposition to vessel wall, early mild inflammatory response surrounding stent struts, typical vascular damage and healing response to acute dilation and a progressive smooth neointimal growth covering stent struts over time. CONCLUSIONS: In this in vivo study of the Minima stent, there was high implant success, predictable re-dilatability to adult diameters and favorable histopathology. Further study is warranted.


Subject(s)
Pulmonary Artery , Stents , Adult , Animals , Aorta , Humans , Infant , Neointima , Pulmonary Artery/diagnostic imaging , Pulmonary Artery/surgery , Swine , Treatment Outcome
SELECTION OF CITATIONS
SEARCH DETAIL
...