Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Cureus ; 15(10): e47047, 2023 Oct.
Article in English | MEDLINE | ID: mdl-38021774

ABSTRACT

Background Deep inspiration breath-hold (DIBH) has been established as a standard technique to reduce cardiac dose. The part of the heart exposed to radiation can be significantly decreased using the DIBH technique during tangential left-sided breast cancer (LSBC) irradiation. Aim The objective of this study was to investigate the intra-fraction breath-hold stability and inter-fraction consistency of patient breath-hold against the threshold as a function of air volumes in the setting of active breathing coordinator (ABC)-based DIBH (ABC-DIBH) treatment to LSBC. Methods A total of 34 patients treated with external beam radiation therapy (EBRT) to the left breast using the ABC-DIBH device were included. The frequency of breath-holds per fraction and the entire course of treatment along with the total treatment time was evaluated for all patients. A prescription dose of either 200 cGy (conventional) or 267 cGy (hypofractionation) was administered during 649 fractions, resulting in a total of 4,601 breath-hold measurements being evaluated. The amplitude of deviation in terms of air volumes between the baseline threshold and the patient-specific measurement (during each breath-hold) per fraction was used to define the DIBH stability. Likewise, the consistency of the breathing amplitudes was used to define the compliance of patient breath-holds throughout the entire treatment period. Positional accuracy was evaluated using orthogonal (portal) images. Results The average number of breath-holds measured over the entire course of treatment for each patient was 144 inspirations (58-351). Similarly, the average number of breath-holds for each fraction during the course of treatment was 11 inspirations (7-21), which included setup imaging and treatment. The total number of breath-holds reduced significantly (p-value < 0.05) with hypofractionation (104 inspirations; range 58-170) as compared to conventional fractionation (145 inspirations; 58-351). The average breath-hold threshold in terms of air volume was 1.41 L (0.6-2.1 L) for all patients. The total treatment time reduced significantly after the third fraction (p-value < 0.05). The average deviation between the measured and baseline threshold breath-holds during the course of treatment was 0.5 L/sec (0.12-1.32 L/sec). The consistency of the breathing amplitudes were maintained within ±0.05 L during the entire treatment for all patients. The average translational shifts measured during setup were 0.28 cm ± 0.3 cm, 0.38 cm ± 0.4 cm, and 0.21 cm ± 0.3 cm in the lateral, longitudinal, and vertical directions, respectively. Conclusion The study has demonstrated the variations in intra-fraction breath-hold stability and inter-fraction breath-hold consistency in terms of air volumes for patients who were treated for LSBC. The frequency of breath-holds was observed to be higher with increased total treatment time for the first few fractions and reduced over the course of treatment.

SELECTION OF CITATIONS
SEARCH DETAIL
...