J Clin Neurosci ; 65: 134-139, 2019 07.
Article in English | MEDLINE | ID: mdl-30852074

ABSTRACT

In 2001, a patient with a deep brain stimulator (DBS) died following treatment with medical diathermy. Manufacturers have since advised against all forms of diathermy except bipolar electrosurgery in DBS patients. This effective ban on monopolar electrosurgery has an impact on the 150,000 patients treated with DBS to date, a number that is set to progressively increase. Analysis of the events, technical specifications, and literature suggests that the original ban was based on extrapolation from medical diathermy to electrosurgery, two very different treatment modalities. This prompted novel work exploring the impact of electrosurgery on DBS systems. Monopolar electrosurgery was employed on an animal cadaveric model with a DBS system paired with a thermocouple at the brain implant site. Prolonged use of monopolar, including at settings higher than normal surgical practice, resulted in a maximum mean temperature increase of only 2.6 °C. Microscopic post-event analysis showed no evidence of thermal injury at the implant site. The implication is that there may be limits within which monopolar electrosurgery use is safe in patients with DBS.


Subject(s)
Diathermy/methods , Electrosurgery/methods , Animals , Brain , Cadaver , Deep Brain Stimulation , Disease Models, Animal , Humans , Sheep