1.
Fresenius J Anal Chem ; 367(1): 86-9, 2000 May.
Article in English | MEDLINE | ID: mdl-11227441

ABSTRACT

The construction and the application of a silica gel based carbon-epoxy indicator electrode for the potentiometric determination of acids and bases are described. The effect of composition of silica gel and carbon-epoxy, slope (mV/pH), linear response (pH range) and the use for acid-base titrations were investigated. The data obtained for the acid-base titrations were compared with those obtained using a glass electrode in the same conditions. The electrode showed a linear response in the pH 2 to 13 range with a slope of -40.5 +/- 0.4 mV/pH (at 25 degrees C) and a response time of less than 15 s. The lifetime of the electrode was higher than one year (over 6000 determinations) with a decrease of only 5% of the initial potentiometric response. The silica gel based carbon-epoxy electrode showed excellent results in the end-point indication potentiometric titrations in determination of acids and bases. The miniaturization of the proposed electrode for flow injection analysis was investigated.