Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters











Database
Language
Publication year range
1.
Tex Heart Inst J ; 41(1): 61-3, 2014 Feb.
Article in English | MEDLINE | ID: mdl-24512403

ABSTRACT

Implantable cardioverter-defibrillators (ICDs) are the standard of care for preventing sudden cardiac death in patients who are predisposed to malignant ventricular arrhythmias. Causes of inappropriate ICD shock include equipment malfunction, improper arrhythmia evaluation, misinterpretation of myopotentials, and electromagnetic interference. As the number of implanted ICDs has increased, other contributors to inappropriate therapy have become known, such as minimal electrical current leaks that mimic ventricular fibrillation. We present the case of a 63-year-old man with a biventricular ICD who received 2 inappropriate shocks, probably attributable to alternating-current leaks in a swimming pool. In addition, we discuss ICD sensitivity and offer recommendations to avoid similar occurrences.


Subject(s)
Defibrillators, Implantable/adverse effects , Electric Countershock/adverse effects , Electric Countershock/instrumentation , Prosthesis Failure , Swimming Pools , Electric Conductivity , Electrophysiologic Techniques, Cardiac , Humans , Male , Middle Aged , Prosthesis Design
SELECTION OF CITATIONS
SEARCH DETAIL