ABSTRACT

The present study was aimed to develop an ultrasensitive technique for electroanalysis of ketamine; a date rape drug. It involved the fabrication of nano-hybrid based electrochemical micro fluidic paper-based analytical device (EµPADs) for electrochemical sensing of ketamine. A paper chip was developed using zeolites nanoflakes and graphene-oxide nanocrystals (Zeo-GO). EµPAD offers many advantages such as facile approach, economical and potential for commercialization. Nanocrystal modified EµPAD showed wide linear range 0.001-5nM/mL and a very low detection limit of 0.001nM/mL. The developed sensor was tested in real time samples like alcoholic and non-alcoholic drinks and found good correlation (99%). The hyphenation of EµPAD integrated with nanocrystalline Zeo-GO for detection of ketamine has immense prospective for field-testing platforms. An extensive development could be made for industrial translation of this fabricated device.


Subject(s)
Anesthetics, Dissociative/analysis , Beverages/analysis , Electrochemical Techniques/instrumentation , Graphite/chemistry , Ketamine/analysis , Point-of-Care Systems , Zeolites/chemistry , Electrodes , Equipment Design , Limit of Detection , Microfluidic Analytical Techniques/instrumentation , Nanoparticles/chemistry , Nanoparticles/ultrastructure , Oxides/chemistry , Paper