J Agric Food Chem ; 46(1): 168-172, 1998 Jan 19.
Article in English | MEDLINE | ID: mdl-10554214

ABSTRACT

This paper describes the development and application of a flow injection analysis system manifold comprising a gas diffusion unit and a potentiometric detector to the determination of free and total SO(2) in white and red wines. A homogeneous crystalline iodide double-membrane tubular electrode was used as detector. SO(2) determination based on the Ripper method was carried out by dosing the iodide formed in the oxidation of SO(2) with iodine, followed by the separation of the formed compound through a diffusion Teflon membrane. The results obtained from the analyses of free and total SO(2) in 30 wine samples showed good agreement between the proposed method and the rapid assay method recommended by the UE and OIV. The relative error deviations of the results obtained by both methods were <6%. This procedure is suitable for samples with approximately 3.2-180 mg L(-)(1) SO(2), performing determinations of 75-100 samples h(-)(1).