Appl Radiat Isot ; 105: 20-25, 2015 Nov.
Article in English | MEDLINE | ID: mdl-26226219

ABSTRACT

The proton beam energy of an isochronous 18MeV cyclotron was determined using a novel version of the stacked copper-foils technique. This simple method used stacked foils of natural copper forming 'thick' targets to produce Zn radioisotopes by the well-documented (p,x) monitor-reactions. Primary beam energy was calculated using the (65)Zn activity vs. depth profile in the target, with the results obtained using (62)Zn and (63)Zn (as comparators) in close agreement. Results from separate measurements using foil thicknesses of 100, 75, 50 or 25µm to form the stacks also concurred closely. Energy was determined by iterative least-squares comparison of the normalized measured activity profile in a target-stack with the equivalent calculated normalized profile, using 'energy' as the regression variable. The technique exploits the uniqueness of the shape of the activity vs. depth profile of the monitor isotope in the target stack for a specified incident energy. The energy using (65)Zn activity profiles and 50-µm foils alone was 18.03±0.02 [SD] MeV (95%CI=17.98-18.08), and 18.06±0.12MeV (95%CI=18.02-18.10; NS) when combining results from all isotopes and foil thicknesses. When the beam energy was re-measured using (65)Zn and 50-µm foils only, following a major upgrade of the ion sources and nonmagnetic beam controls the results were 18.11±0.05MeV (95%CI=18.00-18.23; NS compared with 'before'). Since measurement of only one Zn monitor isotope is required to determine the normalized activity profile this indirect yet precise technique does not require a direct beam-current measurement or a gamma-spectroscopy efficiency calibrated with standard sources, though a characteristic photopeak must be identified. It has some advantages over published methods using the ratio of cross sections of monitor reactions, including the ability to determine energies across a broader range and without need for customized beam degraders.


Subject(s)
Cyclotrons/statistics & numerical data , Protons , Zinc Radioisotopes , Copper , Humans , Radiometry/statistics & numerical data , Zinc Radioisotopes/analysis
2.
Australas Phys Eng Sci Med ; 32(2): 92-7, 2009 Jun.
Article in English | MEDLINE | ID: mdl-19623860

ABSTRACT

A simple and rapid technique to measure the proton beam energy in the external beam line of a medical cyclotron has been examined. A stack of 0.1 mm thick high purity copper (Cu) foils was bombarded and the relative activity of 65Zn produced in each foil was compared to a computational model that predicted activity, based on proton stopping power, reaction cross-sectional data, and beam energy. In the model, the beam energy was altered iteratively until the best match between computed and measured relative activities of the stack of disks was obtained. The main advantage of this method is that it does not require the comparison of the activities of different isotopes of zinc arising from (p, xn) reactions in the Cu, which would require the gamma photon detector being calibrated for different energy responses. Using this technique the proton beam energy of a nominally 18 MeV standard isochronous medical cyclotron was measured as 17.49 +/- 0.04 (SD) MeV, with a precision of 0.2% CV.


Subject(s)
Cyclotrons/standards , Protons , Radiometry/methods , Copper , Reference Standards , Zinc