Phys Med ; 32(1): 237-41, 2016 Jan.
Article in English | MEDLINE | ID: mdl-26508017

ABSTRACT

PURPOSE: Pregnancy during radiosurgery is extremely rare in clinical practice. We report fetal dose results during CyberKnife radiosurgery for a brain tumor in pregnancy. METHODS AND MATERIALS: A 26 year old pregnant woman with a rapidly growing deep-seated grade-III glioma was treated during the third trimester of gestation using CyberKnife. Ultrasound imaging was used to determine the position of the embryo prior to treatment. A dose of 1400 cGy was prescribed aiming to control tumor growth until delivery of the child. Prior to radiosurgery, the treatment was simulated on an anthropomorphic phantom. Radiation dose to the embryo was measured using a Farmer chamber and EBT3 films. RESULTS: Fetal doses of 4.4 cGy and 4.1 cGy were measured for the embryo's head and legs, lying at 56 cm and 72 cm from the isocenter, respectively, using the Farmer chamber situated at 8.5 cm depth beneath the phantom surface. Dose results of 4.4 cGy, 3.5 cGy and 2.0 cGy were measured with the films situated at depths of 6.5 cm, 9.5 cm and 14.5 cm, respectively. An average dose of 4.2 cGy to the fetus was derived from the above values. A corresponding dose of 3.2 cGy was also calculated based on results obtained using EBT3 films situated upon the patient skin. CONCLUSIONS: The measured fetal doses are below the threshold of 10 cGy for congenital malformations, mental and growth retardation effects. The radiogenic cancer risk to the live-born embryo was estimated less than 0.3% over the normal incidence. The treatment was administered successfully, allowing the patient to deliver a healthy child.


Subject(s)
Brain Neoplasms/surgery , Fetus/radiation effects , Glioma/surgery , Pregnancy Complications, Neoplastic/surgery , Radiometry/methods , Radiosurgery/methods , Adult , Anthropometry , Calibration , Female , Humans , Maternal Exposure , Phantoms, Imaging , Pregnancy , Radiation Dosage , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Reproducibility of Results , Treatment Outcome
SELECTION OF CITATIONS
SEARCH DETAIL
...