Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
J Synchrotron Radiat ; 26(Pt 6): 1996-2012, 2019 Nov 01.
Article in English | MEDLINE | ID: mdl-31721745

ABSTRACT

An alternative approach to hard-X-ray photoelectron spectroscopy (HAXPES) has been established. The instrumental key feature is an increase of the dimensionality of the recording scheme from 2D to 3D. A high-energy momentum microscope detects electrons with initial kinetic energies up to 8 keV with a k-resolution of 0.025 Å-1, equivalent to an angular resolution of 0.034°. A special objective lens with k-space acceptance up to 25 Å-1 allows for simultaneous full-field imaging of many Brillouin zones. Combined with time-of-flight (ToF) parallel energy recording this yields maximum parallelization. Thanks to the high brilliance (1013 hν s-1 in a spot of <20 µm diameter) of beamline P22 at PETRA III (Hamburg, Germany), the microscope set a benchmark in HAXPES recording speed, i.e. several million counts per second for core-level signals and one million for d-bands of transition metals. The concept of tomographic k-space mapping established using soft X-rays works equally well in the hard X-ray range. Sharp valence band k-patterns of Re, collected at an excitation energy of 6 keV, correspond to direct transitions to the 28th repeated Brillouin zone. Measured total energy resolutions (photon bandwidth plus ToF-resolution) are 62 meV and 180 meV FWHM at 5.977 keV for monochromator crystals Si(333) and Si(311) and 450 meV at 4.0 keV for Si(111). Hard X-ray photoelectron diffraction (hXPD) patterns with rich fine structure are recorded within minutes. The short photoelectron wavelength (10% of the interatomic distance) `amplifies' phase differences, making full-field hXPD a sensitive structural tool.

SELECTION OF CITATIONS
SEARCH DETAIL
...