Mater Sci Eng C Mater Biol Appl ; 78: 124-129, 2017 Sep 01.
Article in English | MEDLINE | ID: mdl-28575963

ABSTRACT

Graphene Oxide-Molecular Imprinted Polymer (GO-MIP) based electrochemical sensor was developed for the first time towards enzyme less determination of glucose. This GO-MIP was obtained from a series of fictionalization, polymerization and template molecule introduction/removal during the synthesizing process. The proposed GO-MIP based electrode showed excellent electrocatalytic activity towards glucose oxidation at optimized conditions and possessing detection limit of 0.1nM with a response time of ~2min. The current response of GO-MIP based glucose sensor was linearly related to the concentration of glucose. The results obtained from the real time usability of electrodes in human blood matches well with commercially available glucose monitors. Further, the reusability of the material is checked up to eight cycles and interference of glucose with ascorbic acid (AA), uric acid (UA) and dopamine (DA) were also studied. The obtained results endorse the promising application of GO-MIP towards superior glucose sensing with long term stability.


Subject(s)
Electrochemical Techniques , Dopamine , Electrodes , Glucose , Graphite , Humans , Oxides , Polymers