Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
J Hosp Infect ; 84(2): 173-7, 2013 Jun.
Article in English | MEDLINE | ID: mdl-23669262

ABSTRACT

BACKGROUND: Disinfection of the intraluminal space of single-lumen polymer tubes can be obtained by ultraviolet C (UVC) light exposure from an external light source. In existing catheters UVC disinfection is hampered by the design of the catheter hub and tube connector. AIM: To demonstrate that it is possible to design a single-lumen catheter with a hub, tube connector and tube parts that can be UVC-disinfected throughout its entire lumen. METHODS: Two single-lumen catheters were designed: one control and one for UVC exposure. They were contaminated with Pseudomonas aeruginosa (10(4)-10(5) cfu/mL) before UVC light exposure, sampling and plate counting. FINDINGS: Two minutes of UVC exposure was sufficient to obtain 4 log10 disinfection for the full-length prototype catheter. This exposure corresponds to ∼40 mJ/cm(2) at the catheter tip and indicates that even shorter exposure times can be achieved. CONCLUSIONS: It is possible to design catheters that can be disinfected throughout the entire lumen. UVC light exposure could be useful as a decontamination method for catheters in clinical use.


Subject(s)
Catheters/microbiology , Disinfection/methods , Ultraviolet Rays , Colony Count, Microbial , Humans , Microbial Viability/radiation effects , Pseudomonas aeruginosa/radiation effects
SELECTION OF CITATIONS
SEARCH DETAIL
...