Ultrasound Med Biol ; 32(9): 1423-9, 2006 Sep.
Article in English | MEDLINE | ID: mdl-16965982

ABSTRACT

Ultrasound (US) physiotherapy as a clinical treatment is extremely common in the Western world. Internationally, regulation to ensure safe application of US physiotherapy by regular calibration ranges from nil to mandatory. The need for a portable power standard (PPS) has been addressed within a European Community (EC)-funded project. This PPS consists of an electrical driver, a set of US transducers and a cavitation detector (CD). Each component has been extensively tested for stability and travel robustness. Transducer output power has been determined with an uncertainty of <3.3% and with a long-term (2-y) output stability of better than 3%. The CD can detect bubble activity for powers above 3 W for a 1-MHz transducer. Travel trials demonstrated the utility of the PPS in practical measurement environments. Deviations in power measurements observed during these trials were mostly acceptable (<10%), although there were also examples of gross differences (>100%). The PPS is now ready to be used to underpin traceable calibration of physiotherapy devices.


Subject(s)
Physical Therapy Modalities/standards , Ultrasonic Therapy/standards , Calibration/standards , Electronics, Medical/instrumentation , Equipment Design , European Union , Humans , Physical Therapy Modalities/instrumentation , Temperature , Transducers , Ultrasonic Therapy/instrumentation