Spinal Cord ; 42(11): 652-4, 2004 Nov.
Article in English | MEDLINE | ID: mdl-15249927

ABSTRACT

STUDY DESIGN: Case report of an unusual use of AMS 800 (American Medical Systems, Inc., Minnetonka, Minnesota) artificial urinary sphincter cuff in a female patient affected by neurogenic urinary incontinence. OBJECTIVE: To describe this rare surgical solution. SETTING: Department of Urology in Italy. METHODS: A 43-year-old woman affected by flaccid paraplegia, acontractile bladder and incompetent bladder neck, underwent an implantation of an artificial urinary sphincter AMS 800. After 7 years, a mechanical failure of the device occurred and pubovaginal sling (PVS) utilizing the cuff of the sphincter was employed due to the poor quality of rectus fascia and the development of previous allergy for some heterologous materials. RESULTS: At 17 months follow-up, the patient is content and able to empty the bladder by clean intermittent self-catheterization (CIC). CONCLUSION: The risk of developing an allergy reaction due to the employment of heterologous materials and the impossibility to use the rectus fascia obliged us to adopt the pre-existent cuff of the artificial urinary sphincter AMS 800.


Subject(s)
Urinary Bladder, Neurogenic/surgery , Urinary Sphincter, Artificial , Urogenital Surgical Procedures/instrumentation , Adult , Female , Humans , Prosthesis Failure , Urinary Catheterization