1.
Nucleus (La Habana) ; (57): 1-4, Jan.-June 2015.
Article in English | LILACS | ID: lil-754870

ABSTRACT

In this paper, we used the Monte Carlo simulations method in a well-type HPGe detector using directly the manufacturer supplied data in order to simulate the effi ciency response at 46.54 keV. The efficiency values were calculated as a function of the fi lling height of the sample into the measurement geometry and results were found in good agreement with experimental data. The main deviations were less than 2.5 % with a mean of 0.9 %, which is totally satisfactory for the purposes of environmental samples measurements. We also present a brief discussion about the response of the detector to different values of its geometric parameters.


En el trabajo se utilizaron los métodos de simulación por Monte Carlo en un detector HPGe tipo pozo, usando directamente los parámetros del fabricante para simular la respuesta en eficiencia a la energía de 46.54 keV. Los valores de efi ciencia se calcularon en función de la altura de la muestra en la geometría de medición y los resultados se correspondieron con los valores experimentales. Las mayores diferencias fueron menores que el 2.5 % con un promedio de 0.9 %, lo que es totalmente satisfactorio para la medición de muestras ambientales. Se presentó una breve discusión sobre la respuesta del detector para diferentes parámetros geométrico.

2.
Appl Radiat Isot ; 97: 59-62, 2015 Mar.
Article in English | MEDLINE | ID: mdl-25544663

ABSTRACT

Monte Carlo efficiency transfer method was used to determine the full energy peak efficiency of a coaxial n-type HPGe detector. The efficiencies calibration curves for three Certificate Reference Materials were determined by efficiency transfer using a (152)Eu reference source. The efficiency values obtained after efficiency transfer were used to calculate the activity concentration of the radionuclides detected in the three materials, which were measured in a low-background gamma spectrometry system. Reported and calculated activity concentration show a good agreement with mean deviations of 5%, which is satisfactory for environmental samples measurement.

SELECTION OF CITATIONS
SEARCH DETAIL
...