Sensors (Basel) ; 14(6): 11260-76, 2014 Jun 24.
Article in English | MEDLINE | ID: mdl-24961217

ABSTRACT

We describe an ultrasonic instrument for continuous real-time analysis of the fractional mixture of a binary gas system. The instrument is particularly well suited to measurement of leaks of a high molecular weight gas into a system that is nominally composed of a single gas. Sensitivity < 5 × 10(-5) is demonstrated to leaks of octaflouropropane (C3F8) coolant into nitrogen during a long duration (18 month) continuous study. The sensitivity of the described measurement system is shown to depend on the difference in molecular masses of the two gases in the mixture. The impact of temperature and pressure variances on the accuracy of the measurement is analysed. Practical considerations for the implementation and deployment of long term, in situ ultrasonic leak detection systems are also described. Although development of the described systems was motivated by the requirements of an evaporative fluorocarbon cooling system, the instrument is applicable to the detection of leaks of many other gases and to processes requiring continuous knowledge of particular binary gas mixture fractions.


Subject(s)
Chemistry Techniques, Analytical/instrumentation , Complex Mixtures/analysis , Gases/analysis , Microchemistry/instrumentation , Ultrasonography/instrumentation , Chemistry Techniques, Analytical/methods , Equipment Design , Equipment Failure Analysis , Microchemistry/methods , Ultrasonography/methods
SELECTION OF CITATIONS