ABSTRACT

Hospital facilities issue numerous risk announcements on corrosion, deformation or premature wearout of medical devices every year. As there is yet little data on the impact of reprocessing on the quality and durability of microsurgical instruments, this paper aims at evaluating the effects of the reprocessing on microsurgical instruments. Material and Methods 22 brand new microsurgical instruments for stapes surgery were being reprocessed 30 times without being used for surgery or other purposes in the interim time. After each reprocessing the instruments were examined macroscopicly and microscopicly. The results were portrayed in a photo documentation and analysed on that basis. Results Almost all devices showed mechanical damage caused by the reprocessing procedure. The increasing deterioration was often associated with missing protective caps. Furthermore contaminations and stains were apparent in several cases. Conclusions The findings illustrate that careful handling of delicate surgical devices during reprocessing is vital. They also highlight problems of protective caps. As an alternative going forward it should be considered to store microsurgical instruments statically in special racks.


Subject(s)
Equipment Failure , Microsurgery/instrumentation , Sterilization , Equipment Contamination , Equipment Safety , Humans , Risk Factors , Stapes Surgery/instrumentation