Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters










Database
Language
Publication year range
1.
J Bone Joint Surg Br ; 73(4): 569-75, 1991 Jul.
Article in English | MEDLINE | ID: mdl-2071637

ABSTRACT

Nine uncemented hip prostheses of modular design were revised because of late infection (2 cases), femoral stem loosening or fracture (2), loosening of threaded cups (3) and protrusion of bipolar cups (2). At surgery we found tissue discoloration and macroscopic corrosion in four of the nine prostheses, located at the head-neck junction. Histological examination in most cases showed extensive necrosis associated with metal particles, indicating metal toxicity. Metallographic examination of the prosthetic heads revealed structural imperfections and inhomogeneity of the metal. We suggest that the crevice between the head and neck is a potential site of corrosion in modular designs.


Subject(s)
Chromium Alloys/adverse effects , Connective Tissue Diseases/pathology , Hip Prosthesis , Prosthesis Design/adverse effects , Prosthesis Failure , Adolescent , Adult , Animals , Connective Tissue Diseases/chemically induced , Connective Tissue Diseases/diagnostic imaging , Corrosion , Female , Fibrosis , Follow-Up Studies , Giant Cells/pathology , Histiocytes/pathology , Humans , Inflammation/pathology , Male , Microscopy, Electron , Middle Aged , Necrosis , Radiography , Reoperation
SELECTION OF CITATIONS