Mikrochim Acta ; 189(7): 243, 2022 06 03.
Article in English | MEDLINE | ID: mdl-35657569

ABSTRACT

The development of the first microfluidic paper-based analytical device (µPAD) for the speciation of inorganic arsenic in environmental aqueous samples as arsenite (As(III)) and arsenate (As(V)) which implements hydride generation on a paper platform is described. The newly developed µPAD has a 3D configuration and uses Au(III) chloride as the detection reagent. Sodium borohydride is used to generate arsine in the device's sample zone by reducing As(III) in the presence of hydrochloric acid or both As(III) and As(V) (total inorganic As) in the presence of sulfuric acid. Arsine then diffuses across a hydrophobic porous polytetrafluoroethylene membrane into the device's detection zone where it reduces Au(III) to Au nanoparticles. This results in a color change which can be related to the concentration of As(III) or total inorganic As (i.e., As(III) and As(V)) concentration. Under optimal conditions, the µPAD is characterized by a limit of detection of 0.43 mg L-1 for total inorganic As (As(III) + As(V)) and 0.41 mg L-1 for As(III) and a linear calibration range in both cases of 1.2-8.0 mg As L-1. The newly developed µPAD-based method was validated by applying it to groundwater and freshwater samples and comparing the results with those obtained by conventional atomic spectrometric techniques.


Subject(s)
Arsenic , Metal Nanoparticles , Arsenic/analysis , Gold , Microfluidics , Water/analysis
SELECTION OF CITATIONS