Rev Sci Instrum ; 87(11): 11E547, 2016 Nov.
Article in English | MEDLINE | ID: mdl-27910649

ABSTRACT

The Thomson scattering diagnostic on MST records both equilibrium and fluctuating electron temperature with a range capability of 10 eV-5 keV. Standard operation with two modified commercial Nd:YAG lasers allows measurements at rates of 1 kHz-25 kHz. Several subsystems of the diagnostic are being improved. The power supplies for the avalanche photodiode detectors (APDs) that record the scattered light are being replaced to improve usability, reliability, and maintainability. Each of the 144 APDs will have an individual rack mounted switching supply, with bias voltage adjustable to match the APD. Long-wavelength filters (1140 nm center, 80 nm bandwidth) have been added to the polychromators to improve capability to resolve non-Maxwellian distributions and to enable directed electron flow measurements. A supercontinuum (SC) pulsed white light source has replaced the tungsten halogen lamp previously used for spectral calibration of the polychromators. The SC source combines substantial brightness produced in nanosecond pulses with a spectrum that covers the entire range of the polychromators.

2.
Rev Sci Instrum ; 87(11): 11E511, 2016 Nov.
Article in English | MEDLINE | ID: mdl-27910553

ABSTRACT

The Thomson scattering diagnostic on the Madison Symmetric Torus (MST) records excessive levels of stray Nd:YAG laser light. Stray light saturates the 1064 nm spectral channel in all polychromators, which prevents absolute electron density measurements via Rayleigh scattering calibration. Furthermore, stray light contaminates adjacent spectral channels for r/a ≥ 0.75, which renders the diagnostic unable to make electron temperature measurements at these radii. In situ measurements of stray light levels during a vacuum vessel vent are used to identify stray light sources and strategies for reduction of stray light levels. Numerical modeling using Zemax OpticStudio supports these measurements. The model of the vacuum vessel and diagnostic includes synthetic collection optics to enable direct comparison of measured and simulated stray light levels. Modeling produces qualitatively similar stray light distributions to MST measurements, and quantifies the mitigation effects of stray light mitigation strategies prior to implementation.

3.
Rev Sci Instrum ; 83(10): 10E324, 2012 Oct.
Article in English | MEDLINE | ID: mdl-23126982

ABSTRACT

Calibration of the Madison Symmetric Torus Thomson scattering system has been refined to improve temperature fluctuation measurements. Multiple avalanche photodiodes have been directly calibrated for use as reference detectors during calibration, improving accuracy and ease of use. From the absolute calibration we calculate corrections to the gain for variation in detector operating temperature. We also measure the spatial uniformity of detector responsivity for several photodiodes, and present a method of accounting for non-uniformity in the calibration process. Finally, the gain and noise enhancement are measured at multiple wavelengths to improve temperature and uncertainty measurements.

4.
Rev Sci Instrum ; 81(10): 10D513, 2010 Oct.
Article in English | MEDLINE | ID: mdl-21033868

ABSTRACT

Two standard commercial flashlamp-pumped Nd:YAG (YAG denotes yttrium aluminum garnet) lasers have been upgraded to "pulse-burst" capability. Each laser produces a burst of up to 15 2 J Q-switched pulses (1064 nm) at repetition rates of 1-12.5 kHz. Variable pulse-width drive (0.15-0.39 ms) of the flashlamps is accomplished by insulated gate bipolar transistor (IGBT) switching of electrolytic capacitor banks. Direct control of the laser Pockels cell drive enables optimal pulse energy extraction, and up to four 2 J laser pulses during one flashlamp pulse. These lasers are used in the Thomson scattering plasma diagnostic system on the MST reversed-field pinch to record the dynamic evolution of the electron temperature profile and temperature fluctuations. To further these investigations, a custom pulse-burst laser system with a maximum pulse repetition rate of 250 kHz is now being commissioned.

5.
Rev Sci Instrum ; 79(10): 10E733, 2008 Oct.
Article in English | MEDLINE | ID: mdl-19044549

ABSTRACT

The multipoint Thomson scattering diagnostic on the Madison Symmetric Torus (MST) is now fully operational with 21 spatial points, which cover the entire minor radius. Four full electron temperature profiles can be obtained during each MST discharge, with a variable delay between each profile. This system overcomes challenges that arise from the unique machine design, location, and plasma characteristics of MST. The machine design limits the maximum porthole diameter to 11.4 cm, requiring a compact, re-entrant, seven element lens for scattered light collection. Limited space near MST necessitates a long beam path for the two Nd:YAG lasers requiring a remote beam line adjustment system to suppress drift in the beam position due to thermal expansion of the building. Due to the remote location of the laser head, substantial design effort was put into the creation of a set of safety interlocks for the laser system. The dynamic nature of MST plasmas and the wide range of operating space require a versatile scattered light detection system consisting of filter polychromators with temperature controlled avalanche photodiode detectors. We also implement an insertable integrating sphere, which travels along the laser beam path through the vacuum vessel, for the alignment of both the fiber optics and the lasers.

6.
Rev Sci Instrum ; 79(10): 10E734, 2008 Oct.
Article in English | MEDLINE | ID: mdl-19044550

ABSTRACT

Detailed calibrations of the Madison Symmetric Torus polychromator Thomson scattering system have been made suitable for electron temperature fluctuation measurements. All calibrations have taken place focusing on accuracy, ease of use and repeatability, and in situ measurements wherever possible. Novel calibration processes have been made possible with an insertable integrating sphere (ISIS), using an avalanche photodiode (APD) as a reference detector and optical parametric oscillator (OPO). Discussed are a novel in situ spatial calibration with the use of the ISIS, the use of an APD as a reference detector to streamline the APD calibration process, a standard dc spectral calibration, and in situ pulsed spectral calibration made possible with a combination of an OPO as a light source, the ISIS, and an APD used as a reference detector. In addition a relative quantum efficiency curve for the APDs is obtained to aid in uncertainty analysis.

7.
Rev Sci Instrum ; 79(10): 10E735, 2008 Oct.
Article in English | MEDLINE | ID: mdl-19044551

ABSTRACT

The Madison Symmetric Torus (MST) presents challenging conditions for Thomson scattering (TS) measurements. The MST plasmas are reversed-field pinches (RFPs) with electron density n(e)<3x10(13) cm(-3), typically 1x10(13) cm(-3). The TS system was designed to measure from 10 eV to 2 keV; however, six polychromators were upgraded from four to eight spectral channels to resolve to 10 keV. There is no diverter or vertical field, so wall interaction results in high background light both from ion and neutral bremsstrahlungs and from line radiation. Also during standard plasmas, the RFP exhibits regular reconnection sawteeth events during which the plasma current, density, and temperature profiles are flattened. These events are of interest both due to the reconnection physics and to their contribution to the MST equilibrium and confinement. These events occur over 100 microS and exhibit large changes in background light and fast changes in temperature. During improved confinement plasmas, there are no sawteeth; the background is low but the temperature can be over an order of magnitude higher. Data analysis of the system has been developed to accommodate both the large dynamic range of the temperature, the fast dynamics, and the fast changing, high amplitude background. Special attention has been paid to the sources of error, in particular, the contribution of the background. A response-function method reduces the measured uncertainty by a factor of 2. Numerical techniques have been developed which are extremely robust. Two methods are used, a conventional chi(2) minimization using a Levenberg-Marquardt algorithm coupled with Monte Carlo modeling for the error bar and a Bayesian statistics method. The Bayesian method computes the probability distribution for the measured photons and electron temperature and this information can be used to ensemble data and will allow future integrated data analysis efforts.