Clin Chem Lab Med ; 2024 Feb 05.
Article in English | MEDLINE | ID: mdl-38353168

ABSTRACT

To be accurate and equivalent among assays, laboratory results should be traceable to higher-order references and their quality should fulfill maximum allowable measurement uncertainty (MU) as defined to fit the intended clinical use. Accordingly, laboratory professionals should estimate and validate MU of performed tests using appropriate analytical performance specifications (APS). Current consensus supports the derivation of APS by using one of the three models established by the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Strategic Conference held in Milan in 2014. It is recognized that some models are better suited for certain measurands than for others and the attention should be primarily directed towards their biological and clinical characteristics. Among others, model 3 should reflect the state of the art of the measurements that can be defined as the best analytical performance that is technically achievable. Taking serum C-reactive protein and ferritin as examples, here we describe the theoretical premises and the experimental protocol to be used to derive APS for MU when a measurand is allocated to this model. Although the model lacks a direct relationship with clinical outcomes, useful information about the in vitro diagnostic medical device performance and the average quality of provided results may be obtained.

4.
Clin Chem Lab Med ; 61(9): 1552-1557, 2023 08 28.
Article in English | MEDLINE | ID: mdl-36773318

ABSTRACT

From External Quality Assessment data, current harmonization of CRP measuring systems appears to be satisfactory, the inter-assay CV being well below 10%. The inter-method variability is even better (close to 3%) when the widely used measuring systems are compared at CRP concentrations employed as cut-off for detecting sub-clinical infection (i.e., 10.0 mg/L) and measurement variability estimated, according to ISO 20914:2019 Technical Specification, from the intermediate within-lab reproducibility of 6-month consecutive measurement data. According to the state-of-the-art model (which is better suited for CRP), the maximum allowable measurement uncertainty (MAU) for CRP measurement on clinical samples with 10.0 mg/L concentrations is 3.76% (desirable quality). As measurement uncertainty (MU) of the only available reference material (ERM-DA474/IFCC) is ∼3%, to fulfil desirable MAU on clinical samples, IVD manufacturers should work to keep the contribution of remaining MU sources (commercial calibrator and intermediate within-lab reproducibility) lower than 2.3%.


Subject(s)
C-Reactive Protein , Humans , Reference Standards , Uncertainty , Reproducibility of Results , Calibration
6.
Clin Chem Lab Med ; 61(2): 213-223, 2023 01 27.
Article in English | MEDLINE | ID: mdl-36282875

ABSTRACT

Laboratories should estimate and validate [using analytical performance specifications (APS)] the measurement uncertainty (MU) of performed tests. It is therefore essential to appropriately define APS for MU, but also to provide a perspective on suitability of the practical application of these APS. In this study, 23 commonly ordered measurands were allocated to the models defined during the 2014 EFLM Strategic Conference to derive APS for MU. Then, we checked if the performance of commercial measuring systems used in our laboratory may achieve them. Most measurands (serum alkaline phosphatase, aspartate aminotransferase, creatine kinase, γ-glutamyltransferase, lactate dehydrogenase, pancreatic amylase, total proteins, immunoglobulin G, A, M, magnesium, urate, and prostate-specific antigen, plasma homocysteine, and blood red and white cells) were allocated to the biological variation (BV) model and desirable APS were defined accordingly (2.65%, 4.75%, 7.25%, 4.45%, 2.60%, 3.15%, 1.30%, 2.20%, 2.50%, 2.95%, 1.44%, 4.16%, 3.40%, 3.52%, 1.55%, and 5.65%, respectively). Desirable APS for serum total cholesterol (3.00%) and urine albumin (9.00%) were derived using outcome-based model. Lacking outcome-based information, serum albumin, high-density lipoprotein cholesterol, triglycerides, and blood platelets were temporarily reallocated to BV model, the corresponding desirable APS being 1.25%, 2.84%, 9.90%, and 4.85%, respectively. A mix between the two previous models was employed for serum digoxin, with a 6.00% desirable APS. In daily practice by using our laboratory systems, 16 tests fulfilled desirable and five minimum APS, while two (serum albumin and plasma homocysteine) exceeded goals, needing improvements.


Subject(s)
Laboratories , Serum Albumin , Male , Humans , Uncertainty , Cholesterol , Homocysteine
7.
Clin Chem ; 68(9): 1202-1210, 2022 09 01.
Article in English | MEDLINE | ID: mdl-35794075

ABSTRACT

BACKGROUND: Serum ferritin is considered a suitable biomarker of iron-related disorders. However, data about the comparability of results among commercial measuring systems (MSs) are contradictory. We performed an intercomparison study aimed at verifying the current interassay variability and its impact on clinical application of the test. Obtaining this information is vital because manufacturers continue to claim calibration alignment to different WHO preparations, which are not related to each other in terms of traceability. METHODS: Four widely used MSs were evaluated. The interassay agreement was verified using 39 human serum pools. The recovery of WHO International Standard (IS) 94/572 (the only reference material available at the time of the study) was evaluated, after assessing the material commutability. Finally, an approach for harmonizing ferritin results was proposed. RESULTS: Highly significant differences (P < 0.00001) among ferritin concentrations assayed by different MSs were detected and the interassay CV (median 22.9%; interquartile range 21.8-25.5) overlapped the desirable intermethod bias (24.6%). IS 94/572 was commutable for use only with Access and Centaur, with Access being the only MS correctly recovering its assigned value. Accordingly, we used regression data against Access to recalibrate MSs, indirectly aligning them to IS 94/572, with a substantial improvement in degree of harmonization and traceability to higher-order reference. CONCLUSIONS: The harmonization among evaluated ferritin MSs is far from optimal, with the implementation of traceability to different WHO ISs being a factor of confusion. A recalibration approach, however, would permit measurement harmonization, allowing the use of common decision thresholds.


Subject(s)
Ferritins , Iron , Biomarkers , Calibration , Humans , Reference Standards
8.
Clin Chem Lab Med ; 59(12): 1914-1920, 2021 11 25.
Article in English | MEDLINE | ID: mdl-34533007

ABSTRACT

Although being the recommended laboratory test to diagnose acute pancreatitis, serum pancreatic lipase (LIP) is among the poorly standardized laboratory tests, and laboratory stakeholders often appear to not take enough care of the quality of its measurements. Here we discuss some important issues that, if not correctly managed and solved, make misdiagnosis of acute pancreatitis by using serum LIP a real possibility. First, the current unavailability of a suitable higher-order reference material to be used as common calibrator should be filled up to definitively improve the inter-method bias. Second, knowledge of the analytical characteristics that may explain the defective performance of LIP assays should be deepened. IVD manufacturers should be more explicit in providing this information, including description of their internal protocol for transferring LIP values from internal references to commercial calibrators. Third, recommended models for accurately estimating measurement uncertainty and reliably defining analytical performance specifications for LIP measurements should be applied. Finally, investments considering alternative options for measuring LIP (e.g., targeted to the development of automated LIP immunoassays) should be warranted. All involved stakeholders (standardization bodies, higher-order reference providers, in vitro diagnostics manufacturers, and laboratory professionals) should contribute to fill the existing gap.


Subject(s)
Lipase , Pancreatic Function Tests , Pancreatitis , Acute Disease , Humans , Laboratories , Lipase/blood , Pancreatitis/diagnosis , Reference Standards
11.
Int J Mol Sci ; 20(4)2019 Feb 19.
Article in English | MEDLINE | ID: mdl-30791380

ABSTRACT

Tumors often show intra-tumor heterogeneity because of genotypic differences between all the cells that compose it and that derive from it. Recent studies have shown significant aspects of neuroblastoma heterogeneity that may affect the diagnostic-therapeutic strategy. Therefore, we developed a laboratory protocol, based on the combination of the advanced dielectrophoresis-based array technology and next-generation sequencing to identify and sort single cells individually and carry out their copy number variants analysis. The aim was to evaluate the cellular heterogeneity, avoiding overestimation or underestimation errors, due to a bulk analysis of the sample. We tested the above-mentioned protocol on two neuroblastoma cell lines, SK-N-BE(2)-C and IMR-32. The presence of several gain or loss chromosomal regions, in both cell lines, shows a high heterogeneity of the copy number variants status of the single tumor cells, even if they belong to an immortalized cell line. This finding confirms that each cell can potentially accumulate different alterations that can modulate its behavior. The laboratory protocol proposed herein provides a tool able to identify prevalent behaviors, and at the same time highlights the presence of particular clusters that deviate from them. Finally, it could be applicable to many other types of cancer.


Subject(s)
DNA Copy Number Variations , Genetic Heterogeneity , Neuroblastoma/genetics , Cell Line, Tumor , Chromosome Aberrations , Computational Biology , Humans , Neuroblastoma/metabolism , Neuroblastoma/pathology , Single-Cell Analysis , Whole Genome Sequencing
12.
Reprod Biol Endocrinol ; 17(1): 6, 2019 Jan 20.
Article in English | MEDLINE | ID: mdl-30660193

ABSTRACT

BACKGROUND: Among the factors causing male infertility, one of the most debated is the exposure to environmental contaminants. Recently, the chemical compound Bisphenol A (BPA) has drawn attention from the reproductive science community, due to its ubiquitous presence in day-to-day life. Its toxic action appears to mainly affect the male reproductive system, directly impacting male fertility. MAIN: The purpose of this review is to investigate current research data on BPA, providing an overview of the findings obtained from studies in animal and human models, as well as on its supposed mechanisms of action. CONCLUSION: A clear understanding of BPA action mechanisms, as well as the presumed risks deriving from its exposure, is becoming crucial to preserve male fertility. The development and validation of methodologies to detect BPA toxic effects on reproductive organs can provide greater awareness of the potential threat that this chemical represents.


Subject(s)
Benzhydryl Compounds/poisoning , Environmental Pollutants/poisoning , Infertility, Male/chemically induced , Phenols/poisoning , Reproduction/drug effects , Animals , Benzhydryl Compounds/analysis , Environmental Monitoring/methods , Environmental Pollutants/analysis , Female , Humans , Infertility, Male/diagnosis , Infertility, Male/physiopathology , Male , Phenols/analysis , Reproduction/physiology