Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters










Database
Language
Publication year range
1.
Appl Opt ; 49(23): 4372-8, 2010 Aug 10.
Article in English | MEDLINE | ID: mdl-20697439

ABSTRACT

The spatial resolution of an x-ray CCD detector was determined from the widths of the tungsten x-ray lines in the spectrum formed by a crystal spectrometer in the 58 to 70 keV energy range. The detector had 20 microm pixel, 1700 by 1200 pixel format, and a CsI x-ray conversion scintillator. The spectral lines from a megavolt x-ray generator were focused on the spectrometer's Rowland circle by a curved transmission crystal. The line shapes were Lorentzian with an average width after removal of the natural and instrumental line widths of 95 microm (4.75 pixels). A high spatial frequency background, primarily resulting from scattered gamma rays, was removed from the spectral image by Fourier analysis. The spectral lines, having low spatial frequency in the direction perpendicular to the dispersion, were enhanced by partially removing the Lorentzian line shape and by fitting Lorentzian curves to broad unresolved spectral features. This demonstrates the ability to improve the spectral resolution of hard x-ray spectra that are recorded by a CCD detector with well-characterized intrinsic spatial resolution.

2.
Appl Opt ; 41(25): 5209-17, 2002 Sep 01.
Article in English | MEDLINE | ID: mdl-12211544

ABSTRACT

The time-dependent response of a 1-mm2 silicon photodiode was characterized by use of pulsed synchrotron radiation in the 4- to 16-nm-wavelength range. Modeling the input radiation pulse and the electrical response of the photodiode allowed the photodiode's capacitance as a function of wavelength and applied bias voltage to be determined. The capacitance was in the 7- to 19-pF range and resulted in response fall times as small as 0.4 ns. The capacitance determined by pulsed x-ray illumination was in good agreement with the capacitance determined by pulsed optical laser illumination. The absolute responsivity was measured by comparison with the responsivity of a calibrated photodiode.

SELECTION OF CITATIONS
SEARCH DETAIL
...