Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters










Database
Language
Publication year range
2.
Proc (Bayl Univ Med Cent) ; 32(4): 502-504, 2019 Oct.
Article in English | MEDLINE | ID: mdl-31656405

ABSTRACT

To verify that temporal artery (TA) temperature measured in the postanesthesia care unit (PACU) in noncardiac surgical patients is a valid reflection of core temperature, a prospective, observational, institutional review board-approved study was conducted in a large, academic tertiary care hospital. The study developed from an initial quality improvement project. A total of 276 patients who had an indwelling bladder catheter as standard of care were enrolled when a research student was available over a 6-month period in 2015. Infrared TA temperature was measured (average of three readings) simultaneously with bladder temperature on PACU arrival. Mean temperature in the bladder and TA groups was >36°C with a clinically negligible difference (0.125°C; 90% confidence interval, 0.059-0.192). Agreement between bladder and TA temperatures, as well as between bladder and last operating room temperatures, was >95% by Bland-Altman analysis. A properly performed TA temperature measure on PACU arrival is an acceptable representation of core temperature for purposes of quality assessment, patient comfort, and regulatory requirements.

SELECTION OF CITATIONS
SEARCH DETAIL
...