1.
Exp Mech ; 54(2): 187-198, 2014 Feb 01.
Article in English | MEDLINE | ID: mdl-24729630

ABSTRACT

To validate models of contact mechanics in low speed structural impact, slender rods were impacted in a drop tower, and measurements of the contact and vibration were compared to analytical and finite element (FE) models. The contact area was recorded using a novel thin-film transfer technique, and the contact duration was measured using electrical continuity. Strain gages recorded the vibratory strain in one rod, and a laser Doppler vibrometer measured speed. The experiment was modeled analytically on a one-dimensional spatial domain using a quasi-static Hertzian contact law and a system of delay differential equations. The three-dimensional FE model used hexahedral elements, a penalty contact algorithm, and explicit time integration. A small submodel taken from the initial global FE model economically refined the analysis in the small contact region. Measured contact areas were within 6% of both models' predictions, peak speeds within 2%, cyclic strains within 12 µÎµ (RMS value), and contact durations within 2 µs. The global FE model and the measurements revealed small disturbances, not predicted by the analytical model, believed to be caused by interactions of the non-planar stress wavefront with the rod's ends. The accuracy of the predictions for this simple test, as well as the versatility of the diagnostic tools, validates the theoretical and computational models, corroborates instrument calibration, and establishes confidence that the same methods may be used in experimental and computational study of contact mechanics during impact of more complicated structures. Recommendations are made for applying the methods to a particular biomechanical problem: the edge-loading of a loose prosthetic hip joint which can lead to premature wear and prosthesis failure.

2.
J Biomed Mater Res B Appl Biomater ; 102(2): 311-21, 2014 Feb.
Article in English | MEDLINE | ID: mdl-23996812

ABSTRACT

This research has developed a novel test method for evaluating the wear resistance of ceramic materials under severe contact stresses simulating edge loading in prosthetic hip bearings. Simply shaped test specimens - a cylinder and a spheroid - were designed as surrogates for an edge-loaded, head/liner implant pair. Equivalency of the simpler specimens was assured in the sense that their theoretical contact dimensions and pressures were identical, according to Hertzian contact theory, to those of the head/liner pair. The surrogates were fabricated in three ceramic materials: Al2 O3 , zirconia-toughened alumina (ZTA), and ZrO2 . They were mated in three different material pairs and reciprocated under a 200 N normal contact force for 1000-2000 cycles, which created small (<1 mm(2) ) wear scars. The three material pairs were ranked by their wear resistance, quantified by the volume of abraded material measured using an interferometer. Similar tests were performed on edge-loaded hip implants in the same material pairs. The surrogates replicated the wear rankings of their full-scale implant counterparts and mimicked their friction force trends. The results show that a proxy test using simple test specimens can validly rank the wear performance of ceramic materials under severe, edge-loading contact stresses, while replicating the beginning stage of edge-loading wear. This simple wear test is therefore potentially useful for screening and ranking new, prospective materials early in their development, to produce optimized candidates for more complicated full-scale hip simulator wear tests.


Subject(s)
Ceramics/chemistry , Hip Prosthesis , Materials Testing , Models, Theoretical , Humans , Stress, Mechanical , Weight-Bearing
3.
J Arthroplasty ; 27(1): 150-2, 2012 Jan.
Article in English | MEDLINE | ID: mdl-21978562

ABSTRACT

Ceramic acetabular liners may exhibit a small, sharp crest-an artifact of discontinuous machining steps--at the junction between the concave spherical surface and the interior edge. On 3 ceramic liners, this crest was found to form a 9° to 11° deviation from tangency. Edge loading wear tests were conducted directly on this crest and on a smoother region of the edge. The crest elicited 2 to 15 times greater volumetric wear on the femoral head. The propensity of the crest to rapidly (<2000 wear cycles) cause elevated wear under low contact force (200 N) suggests that the crest artifact of prevailing machining protocols might be a root cause of stripe wear and squeaking in ceramic acetabular bearings.


Subject(s)
Ceramics , Hip Prosthesis , Materials Testing , Prosthesis Failure , Acetabulum , Prosthesis Design , Surface Properties , Time Factors
4.
J Biomech ; 44(16): 2802-8, 2011 Nov 10.
Article in English | MEDLINE | ID: mdl-21962465

ABSTRACT

The components of prosthetic hip bearings may experience in-vivo subluxation and edge loading on the acetabular socket as a result of joint laxity, causing abnormally high, damaging contact stresses. In this research, edge-loaded contact of prosthetic hips is examined analytically and experimentally in the most commonly used categories of material pairs. In edge-loaded ceramic-on-ceramic hips, the Hertzian contact theory yields accurate (conservatively, <10% error) predictions of the contact dimensions. Moreover, the Hertzian theory successfully captures slope and curvature trends in the dependence of contact patch geometry on the applied load. In an edge-loaded ceramic-on-metal pair, a similar degree of accuracy is observed in the contact patch length; however, the contact width is less accurately predicted due to the onset of subsurface plasticity, which is predicted for loads >400N. The Hertzian contact theory is shown to be ill-suited to edge-loaded ceramic-on-polyethylene pairs due to polyethylene's nonlinear material behavior. This work elucidates the methods and the accuracy of applying classical contact theory to edge-loaded hip bearings. The results help to define the applicability of the Hertzian theory to the design of new components and materials to better resist severe edge loading contact stresses.


Subject(s)
Hip Prosthesis , Materials Testing , Models, Theoretical , Weight-Bearing , Ceramics , Humans , Polyethylene
5.
J Tribol ; 133(2): 2455021-245026, 2011 Apr 01.
Article in English | MEDLINE | ID: mdl-23970811

ABSTRACT

Laboratory testing of contact phenomena can be prohibitively expensive if the interacting bodies are geometrically complicated. This work demonstrates means to mitigate such problems by exploiting the established observation that two geometrically dissimilar contact pairs may exhibit the same contact mechanics. Specific formulas are derived that allow a complicated Hertzian contact pair to be replaced with an inexpensively manufactured and more easily fixtured surrogate pair, consisting of a plane and a spheroid, which has the same (to second-order accuracy) contact area and pressure distribution as the original complicated geometry. This observation is elucidated by using direct tensor notation to review a key assertion in Hertzian theory; namely, geometrically complicated contacting surfaces can be described to second-order accuracy as contacting ellipsoids. The surrogate spheroid geometry is found via spectral decomposition of the original pair's combined Hessian tensor. Some numerical examples using free-form surfaces illustrate the theory, and a laboratory test validates the theory under a common scenario of normally compressed convex surfaces. This theory for a Hertzian contact substitution may be useful in simplifying the contact, wear, or impact testing of complicated components or of their constituent materials.

6.
IEEE Trans Vis Comput Graph ; 11(5): 508-18, 2005.
Article in English | MEDLINE | ID: mdl-16144248

ABSTRACT

Huge salt formations, trapping large untapped oil and gas reservoirs, lie in the deepwater region of the Gulf of Mexico. Drilling in this region is high-risk and drilling failures have led to well abandonments, with each costing tens of millions of dollars. Salt tectonics plays a central role in these failures. To explore the geomechanical interactions between salt and the surrounding sand and shale formations, scientists have simulated the stresses in and around salt diapirs in the Gulf of Mexico using nonlinear finite element geomechanical modeling. In this paper, we describe novel techniques developed to visualize the simulated subsurface stress field. We present an adaptation of the Mohr diagram, a traditional paper-and-pencil graphical method long used by the material mechanics community for estimating coordinate transformations for stress tensors, as a new tensor glyph for dynamically exploring tensor variables within three-dimensional finite element models. This interactive glyph can be used as either a probe or a filter through brushing and linking.


Subject(s)
Algorithms , Computer Graphics , Geologic Sediments/analysis , Geology/methods , Image Interpretation, Computer-Assisted/methods , Information Storage and Retrieval/methods , User-Computer Interface , Data Display , Geologic Sediments/classification , Image Enhancement/methods , Oceans and Seas , Stress, Mechanical