1.
Tex Heart Inst J ; 17(3): 223-7, 1990.
Article in English | MEDLINE | ID: mdl-15227175

ABSTRACT

From August 1971 through November 1972, we implanted 62 Model 2 DeBakey-Surgitool aortic valve prostheses in 62 patients, 4 of whom later had clinically asymptomatic strut fractures. In 1 case, the patient died suddenly, and autopsy revealed detachment of the ball-cage; in each of the other 3 cases, fractures of 2 struts close to the base of the prosthesis were diagnosed fluoroscopically, and the patients underwent successful reoperation. The interval between implantation and reoperation ranged from 11 months to 16 years, 9 months. In 1 patient, retrospective study of chest radiographs revealed that the fracture had been present for 2(1/2) years. Larger valves (>/= A6) were affected significantly more often than smaller ones. We performed metallurgic analysis of 1 prosthesis: results revealed strut wear from fatigue cracking and secondary abrasion. Strut fracture was also promoted by suspension of the cage at right angles to the prosthetic ring and by use of a pyrolytic carbon ball in a titanium cage (i.e., an occluder harder than its holder). Patients with DeBakey-Surgitool aortic valve prostheses should undergo annual radiologic examinations to enable early detection of strut fractures. Prophylactic valve replacement is not indicated.