Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 20 de 23
Filter
1.
Med Phys ; 44(11): 5795-5813, 2017 Nov.
Article in English | MEDLINE | ID: mdl-28801918

ABSTRACT

PURPOSE: In order to mitigate motion-induced artifacts, several motion compensation (MoCo) methods have been developed, which are either able to (a) compensate for severe artifacts, but utilize the data for the reconstruction of several cardiac phases, or (b) improve image quality of a single reconstruction with only moderate motion artifacts. We propose a method combining both benefits: dose efficiency by utilizing only the data needed for a single short-scan reconstruction while still being able to compensate for severe artifacts. METHODS: We introduce a MoCo method, which we call PAMoCo, to improve the visualization of the coronary arteries of a standard coronary CT angiography exam by reducing motion artifacts. As a first step, we segment a region of interest covering a chosen coronary artery. We subdivide a volume covering the whole heart into several stacks, which are sub-volumes, reconstructed from phase-correlated short-scan data acquired during different heart cycles. Motion-compensated reconstruction is performed for each stack separately, based on partial angle reconstructions, which are derived by dividing the data corresponding to the segmented stack volume into several double-overlapping sectors. We model motion along the coronary artery center line obtained from segmentation and the temporal dimension by a low-degree polynomial and create a dense 3D motion vector field (MVF). The parameters defining the MVF are estimated by optimizing an image artifact measuring cost function and we employ a semi-global optimization routine by re-initializing the optimization multiple times. The algorithm was evaluated on the basis of a phantom measurement and clinical data. For the phantom measurement an artificial vessel equipped with calcified lesions mounted on a moving robot arm was measured, where typical coronary artery motion patterns for 70 bpm and 90 bpm have been applied. For analysis, we calculated the calcified volume V inside an ROI and measured the maximum vessel diameter d based on cross-sectional views to compare the performances of standard reconstructions obtained via filtered backprojection (FBP) and PAMoCo reconstructions between 20% and 80% of the cardiac cycle. Further, the new algorithm was applied to six clinical cases of patients with heart rates between 50 bpm and 74 bpm. Standard FBP, PAMoCo reconstructions were performed and compared to best phase FBP reconstructions and another MoCo algorithm, which is based on motion artifact metrics (MAM), via visual inspection. RESULTS: In case of the phantom measurement we found a strong dependence of V and d on the cardiac phase in case of the FBP reconstructions. When applying PAMoCo, V and d became almost constant due to a better discrimination from calcium to vessel and water background and values close to the ground truth have been derived. In the clinical study we chose reconstructions showing strong motion artifacts and observed a substantially improved delineation of the coronary arteries in PAMoCo reconstructions compared to the standard FBP reconstructions and also the MAM images, confirming the findings of the phantom measurement. CONCLUSIONS: Due to the fast reconstruction of PAMoCo images and the introduction of a new motion model, we were able to re-initialize the optimization routine at pre-selected parameter sets and thereby increase the potential of the MAM algorithm. From the phantom measurement we conclude that PAMoCo performed almost equally well in all cardiac phases and suggest applying the PAMoCo algorithm for single source systems in case of patients with high or irregular heart rates.


Subject(s)
Artifacts , Coronary Vessels/diagnostic imaging , Image Processing, Computer-Assisted/methods , Movement , Tomography, X-Ray Computed , Humans , Phantoms, Imaging
2.
Med Phys ; 40(3): 031112, 2013 Mar.
Article in English | MEDLINE | ID: mdl-23464292

ABSTRACT

PURPOSE: The temporal resolution of a given image in cardiac computed tomography (CT) has so far mostly been determined from the amount of CT data employed for the reconstruction of that image. The purpose of this paper is to examine the applicability of such measures to the newly introduced modality of dual-source CT as well as to methods aiming to provide improved temporal resolution by means of an advanced image reconstruction algorithm. METHODS: To provide a solid base for the examinations described in this paper, an extensive review of temporal resolution in conventional single-source CT is given first. Two different measures for assessing temporal resolution with respect to the amount of data involved are introduced, namely, either taking the full width at half maximum of the respective data weighting function (FWHM-TR) or the total width of the weighting function (total TR) as a base of the assessment. Image reconstruction using both a direct fan-beam filtered backprojection with Parker weighting as well as using a parallel-beam rebinning step are considered. The theory of assessing temporal resolution by means of the data involved is then extended to dual-source CT. Finally, three different advanced iterative reconstruction methods that all use the same input data are compared with respect to the resulting motion artifact level. For brevity and simplicity, the examinations are limited to two-dimensional data acquisition and reconstruction. However, all results and conclusions presented in this paper are also directly applicable to both circular and helical cone-beam CT. RESULTS: While the concept of total TR can directly be applied to dual-source CT, the definition of the FWHM of a weighting function needs to be slightly extended to be applicable to this modality. The three different advanced iterative reconstruction methods examined in this paper result in significantly different images with respect to their motion artifact level, despite exactly the same amount of data being used in the reconstruction process. CONCLUSIONS: The concept of assessing temporal resolution by means of the data employed for reconstruction can nicely be extended from single-source to dual-source CT. However, for advanced (possibly nonlinear iterative) reconstruction algorithms the examined approach fails to deliver accurate results. New methods and measures to assess the temporal resolution of CT images need to be developed to be able to accurately compare the performance of such algorithms.


Subject(s)
Artifacts , Heart/diagnostic imaging , Heart/physiology , Movement , Tomography, X-Ray Computed/methods , Humans , Image Processing, Computer-Assisted , Phantoms, Imaging
3.
Med Phys ; 40(3): 031901, 2013 Mar.
Article in English | MEDLINE | ID: mdl-23464316

ABSTRACT

PURPOSE: Research in image reconstruction for cardiac CT aims at using motion correction algorithms to improve the image quality of the coronary arteries. The key to those algorithms is motion estimation, which is currently based on 3-D/3-D registration to align the structures of interest in images acquired in multiple heart phases. The need for an extended scan data range covering several heart phases is critical in terms of radiation dose to the patient and limits the clinical potential of the method. Furthermore, literature reports only slight quality improvements of the motion corrected images when compared to the most quiet phase (best-phase) that was actually used for motion estimation. In this paper a motion estimation algorithm is proposed which does not require an extended scan range but works with a short scan data interval, and which markedly improves the best-phase image quality. METHODS: Motion estimation is based on the definition of motion artifact metrics (MAM) to quantify motion artifacts in a 3-D reconstructed image volume. The authors use two different MAMs, entropy, and positivity. By adjusting the motion field parameters, the MAM of the resulting motion-compensated reconstruction is optimized using a gradient descent procedure. In this way motion artifacts are minimized. For a fast and practical implementation, only analytical methods are used for motion estimation and compensation. Both the MAM-optimization and a 3-D/3-D registration-based motion estimation algorithm were investigated by means of a computer-simulated vessel with a cardiac motion profile. Image quality was evaluated using normalized cross-correlation (NCC) with the ground truth template and root-mean-square deviation (RMSD). Four coronary CT angiography patient cases were reconstructed to evaluate the clinical performance of the proposed method. RESULTS: For the MAM-approach, the best-phase image quality could be improved for all investigated heart phases, with a maximum improvement of the NCC value by 100% and of the RMSD value by 81%. The corresponding maximum improvements for the registration-based approach were 20% and 40%. In phases with very rapid motion the registration-based algorithm obtained better image quality, while the image quality of the MAM algorithm was superior in phases with less motion. The image quality improvement of the MAM optimization was visually confirmed for the different clinical cases. CONCLUSIONS: The proposed method allows a software-based best-phase image quality improvement in coronary CT angiography. A short scan data interval at the target heart phase is sufficient, no additional scan data in other cardiac phases are required. The algorithm is therefore directly applicable to any standard cardiac CT acquisition protocol.


Subject(s)
Heart/diagnostic imaging , Imaging, Three-Dimensional/methods , Movement , Tomography, X-Ray Computed/methods , Algorithms , Artifacts , Humans , Phantoms, Imaging
4.
Eur J Radiol ; 82(2): 270-4, 2013 Feb.
Article in English | MEDLINE | ID: mdl-23137919

ABSTRACT

OBJECTIVE: To evaluate the effect of a temporal resolution improvement method (TRIM) for cardiac CT on diagnostic image quality for coronary artery assessment. MATERIALS AND METHODS: The TRIM-algorithm employs an iterative approach to reconstruct images from less than 180° of projections and uses a histogram constraint to prevent the occurrence of limited-angle artifacts. This algorithm was applied in 11 obese patients (7 men, 67.2 ± 9.8 years) who had undergone second generation dual-source cardiac CT with 120 kV, 175-426 mAs, and 500 ms gantry rotation. All data were reconstructed with a temporal resolution of 250 ms using traditional filtered-back projection (FBP) and of 200 ms using the TRIM-algorithm. Contrast attenuation and contrast-to-noise-ratio (CNR) were measured in the ascending aorta. The presence and severity of coronary motion artifacts was rated on a 4-point Likert scale. RESULTS: All scans were considered of diagnostic quality. Mean BMI was 36 ± 3.6 kg/m(2). Average heart rate was 60 ± 9 bpm. Mean effective dose was 13.5 ± 4.6 mSv. When comparing FBP- and TRIM reconstructed series, the attenuation within the ascending aorta (392 ± 70.7 vs. 396.8 ± 70.1 HU, p>0.05) and CNR (13.2 ± 3.2 vs. 11.7 ± 3.1, p>0.05) were not significantly different. A total of 110 coronary segments were evaluated. All studies were deemed diagnostic; however, there was a significant (p<0.05) difference in the severity score distribution of coronary motion artifacts between FBP (median=2.5) and TRIM (median=2.0) reconstructions. CONCLUSION: The algorithm evaluated here delivers diagnostic imaging quality of the coronary arteries despite 500 ms gantry rotation. Possible applications include improvement of cardiac imaging on slower gantry rotation systems or mitigation of the trade-off between temporal resolution and CNR in obese patients.


Subject(s)
Algorithms , Coronary Angiography/methods , Coronary Artery Disease/diagnostic imaging , Radiographic Image Enhancement/methods , Radiographic Image Interpretation, Computer-Assisted/methods , Tomography, X-Ray Computed/methods , Aged , Female , Humans , Male , Pilot Projects , Reproducibility of Results , Sensitivity and Specificity
5.
J Thorac Imaging ; 25(2): 100-11, 2010 May.
Article in English | MEDLINE | ID: mdl-20463530

ABSTRACT

Multidetector row computed tomography (MDCT) has been used for cardiothoracic imaging since the advent of 4-slice computed tomography (CT) in 1999. Available since 2004, 64-slice CT systems are currently considered a prerequisite for successfully integrating cardiothoracic CT into routine clinical algorithms. Developments are ongoing that aim to solve the remaining challenges of cardiothoracic CT. In this review article, we focus on 3 aspects that have significantly influenced the design of newer CT scanners. (1) A faster scan speed enables coverage of the cardiothoracic anatomy in shorter scan times, which is beneficial for patients with limited ability to cooperate and opens new clinical possibilities. A coveted goal for cardiac CT is a "snapshot image" of the entire heart in 1 cardiac cycle, which can be obtained by using MDCT systems with area detectors or dual-source CT systems with electrocardiogram (ECG)-triggered high-pitch spiral. (2) Techniques to reduce the radiation dose to the patient, such as ECG-controlled dose modulation, ECG-triggered sequential CT, low kV scanning, and iterative reconstruction, have gained considerable attention as a consequence of the ongoing discussion of radiation exposure by CT. (3) New developments aim to establish a role for CT in the functional imaging of the lung and of the heart beyond the mere visualization of anatomy. Evaluation of the first-pass enhancement of the myocardium, with single-energy or with dual-energy data acquisition, and cardiac perfusion examinations may have the potential to enhance the application spectrum of cardiac CT by providing the means to determine the hemodynamic relevance of coronary artery stenosis.


Subject(s)
Cardiovascular Diseases/diagnostic imaging , Radiography, Thoracic/methods , Thoracic Diseases/diagnostic imaging , Tomography Scanners, X-Ray Computed , Tomography, X-Ray Computed/methods , Contrast Media , Electrocardiography , Equipment Design , Humans , Radiation Dosage , Radiation Protection/methods , Radiographic Image Interpretation, Computer-Assisted , Time Factors
6.
J Cardiovasc Comput Tomogr ; 3(3): 143-52, 2009.
Article in English | MEDLINE | ID: mdl-19527890

ABSTRACT

In this article, we review the current status and discuss potential further improvements and limitations of system parameters relevant for cardiac CT, in particular spatial resolution, temporal resolution, and volume coverage.


Subject(s)
Cardiovascular Diseases/diagnosis , Tomography, X-Ray Computed , Cardiovascular Diseases/diagnostic imaging , Humans , Sensitivity and Specificity , Tomography, X-Ray Computed/methods
7.
J Cardiovasc Comput Tomogr ; 3(2): 117-21, 2009.
Article in English | MEDLINE | ID: mdl-19332343

ABSTRACT

Coronary CT angiography allows high-quality imaging of the coronary arteries when state-of-the-art CT systems are used. However, radiation exposure has been a concern. We describe a new scan mode that uses a very high-pitch spiral acquisition, "Flash Spiral," which has been developed specifically for low-dose imaging with dual-source CT. The scan mode uses a pitch of 3.2 to acquire a spiral CT data set, while covering the entire volume of the heart in one cardiac cycle. Data acquisition is prospectively triggered by the electrocardiogram and starts in late systole to be completed within one cardiac cycle. Images are reconstructed with a temporal resolution that corresponds to one-quarter of the gantry rotation time. Throughout the data set, subsequent images are reconstructed at later time instants in the cardiac cycle. In a patient with a heart rate of 49 beats/min, the Flash Spiral scan mode was used with a first-generation dual-source CT system and allowed artifact-free visualization of the coronary arteries with a radiation exposure of 1.7 mSv for a 12-cm scan range at 120 kVp tube voltage.


Subject(s)
Coronary Angiography/methods , Coronary Stenosis/diagnostic imaging , Tomography, Spiral Computed , Coronary Angiography/adverse effects , Electrocardiography , Humans , Male , Middle Aged , Predictive Value of Tests , Radiation Dosage , Radiation Injuries/etiology , Radiation Injuries/prevention & control , Radiographic Image Interpretation, Computer-Assisted
8.
Med Phys ; 36(12): 5641-53, 2009 Dec.
Article in English | MEDLINE | ID: mdl-20095277

ABSTRACT

PURPOSE: To present the theory for image reconstruction of a high-pitch, high-temporal-resolution spiral scan mode for dual-source CT (DSCT) and evaluate its image quality and dose. METHODS: With the use of two x-ray sources and two data acquisition systems, spiral CT exams having a nominal temporal resolution per image of up to one-quarter of the gantry rotation time can be acquired using pitch values up to 3.2. The scan field of view (SFOV) for this mode, however, is limited to the SFOV of the second detector as a maximum, depending on the pitch. Spatial and low contrast resolution, image uniformity and noise, CT number accuracy and linearity, and radiation dose were assessed using the ACR CT accreditation phantom, a 30 cm diameter cylindrical water phantom or a 32 cm diameter cylindrical PMMA CTDI phantom. Slice sensitivity profiles (SSPs) were measured for different nominal slice thicknesses, and an anthropomorphic phantom was used to assess image artifacts. Results were compared between single-source scans at pitch = 1.0 and dual-source scans at pitch = 3.2. In addition, image quality and temporal resolution of an ECG-triggered version of the DSCT high-pitch spiral scan mode were evaluated with a moving coronary artery phantom, and radiation dose was assessed in comparison with other existing cardiac scan techniques. RESULTS: No significant differences in quantitative measures of image quality were found between single-source scans at pitch = 1.0 and dual-source scans at pitch = 3.2 for spatial and low contrast resolution, CT number accuracy and linearity, SSPs, image uniformity, and noise. The pitch value (1.6 pitch 3.2) had only a minor impact on radiation dose and image noise when the effective tube current time product (mA s/pitch) was kept constant. However, while not severe, artifacts were found to be more prevalent for the dual-source pitch = 3.2 scan mode when structures varied markedly along the z axis, particularly for head scans. Images of the moving coronary artery phantom acquired with the ECG-triggered high-pitch scan mode were visually free from motion artifacts at heart rates of 60 and 70 bpm. However, image quality started to deteriorate for higher heart rates. At equivalent image quality, the ECG-triggered high-pitch scan mode demonstrated lower radiation dose than other cardiac scan techniques on the same DSCT equipment (25% and 60% dose reduction compared to ECG-triggered sequential step-and-shoot and ECG-gated spiral with x-ray pulsing). CONCLUSIONS: A high-pitch (up to pitch = 3.2), high-temporal-resolution (up to 75 ms) dual-source CT scan mode produced equivalent image quality relative to single-source scans using a more typical pitch value (pitch = 1.0). The resultant reduction in the overall acquisition time may offer clinical advantage for cardiovascular, trauma, and pediatric CT applications. In addition, ECG-triggered high-pitch scanning may be useful as an alternative to ECG-triggered sequential scanning for patients with low to moderate heart rates up to 70 bpm, with the potential to scan the heart within one heart beat at reduced radiation dose.


Subject(s)
Radiographic Image Interpretation, Computer-Assisted/methods , Tomography, Spiral Computed/methods , Artifacts , Linear Models , Phantoms, Imaging , Radiation Dosage , Time Factors
9.
Med Phys ; 36(12): 5683-94, 2009 Dec.
Article in English | MEDLINE | ID: mdl-20095281

ABSTRACT

PURPOSE: Cardiac CT achieves its high temporal resolution by lowering the scan range from 2pi to pi plus fan angle (partial scan). This, however, introduces CT-value variations, depending on the angular position of the pi range. These partial scan artifacts are of the order of a few HU and prevent the quantitative evaluation of perfusion measurements. The authors present the new algorithm partial scan artifact reduction (PSAR) that corrects a dynamic phase-correlated scan without a priori information. METHODS: In general, a full scan does not suffer from partial scan artifacts since all projections in [0, 2pi] contribute to the data. To maintain the optimum temporal resolution and the phase correlation, PSAR creates an artificial full scan pn(AF) by projectionwise averaging a set of neighboring partial scans pn(P) from the same perfusion examination (typically N approximately 30 phase-correlated partial scans distributed over 20 s and n = 1, ..., N). Corresponding to the angular range of each partial scan, the authors extract virtual partial scans pn(V) from the artificial full scan pn(AF). A standard reconstruction yields the corresponding images fn(P), fn(AF), and fn(V). Subtracting the virtual partial scan image fn(V) from the artificial full scan image fn(AF) yields an artifact image that can be used to correct the original partial scan image: fn(C) = fn(P) - fn(V) + fn(AF), where fn(C) is the corrected image. RESULTS: The authors evaluated the effects of scattered radiation on the partial scan artifacts using simulated and measured water phantoms and found a strong correlation. The PSAR algorithm has been validated with a simulated semianthropomorphic heart phantom and with measurements of a dynamic biological perfusion phantom. For the stationary phantoms, real full scans have been performed to provide theoretical reference values. The improvement in the root mean square errors between the full and the partial scans with respect to the errors between the full and the corrected scans is up to 54% for the simulations and 90% for the measurements. CONCLUSIONS: The phase-correlated data now appear accurate enough for a quantitative analysis of cardiac perfusion.


Subject(s)
Artifacts , Myocardial Perfusion Imaging/methods , Tomography, X-Ray Computed/methods , Heart/diagnostic imaging , Phantoms, Imaging , Water
10.
Eur J Radiol ; 68(3): 362-8, 2008 Dec.
Article in English | MEDLINE | ID: mdl-18842371

ABSTRACT

During the past years, multi-detector row CT (MDCT) has evolved into clinical practice with a rapid increase of the number of detector slices. Today's 64 slice CT systems allow whole-body examinations with sub-millimeter resolution in short scan times. As an alternative to adding even more detector slices, we describe the system concept and design of a CT scanner with two X-ray tubes and two detectors (mounted on a CT gantry with a mechanical offset of 90 degrees) that has the potential to overcome limitations of conventional MDCT systems, such as temporal resolution for cardiac imaging. A dual source CT (DSCT) scanner provides temporal resolution equivalent to a quarter of the gantry rotation time, independent of the patient's heart rate (83 ms at 0.33 s rotation time). In addition to the benefits for cardiac scanning, it allows to go beyond conventional CT imaging by obtaining dual energy information if the two tubes are operated at different voltages. Furthermore, we discuss how both acquisition systems can be used to add the power reserve of two X-ray tubes for long scan ranges and obese patients. Finally, future advances of DSCT are highlighted.


Subject(s)
Radiographic Image Enhancement/instrumentation , Radiographic Image Enhancement/methods , Tomography, X-Ray Computed/instrumentation , Tomography, X-Ray Computed/methods , Equipment Design , Equipment Failure Analysis , Humans , Technology Assessment, Biomedical , Tomography, X-Ray Computed/trends
11.
Med Phys ; 35(2): 764-8, 2008 Feb.
Article in English | MEDLINE | ID: mdl-18383698

ABSTRACT

This work assesses the temporal resolution of dual-source computed tomography (CT) in a visually intuitive manner. Exploiting the principles of ring artifact creation, a phantom containing a highly attenuating delta function signal (a wire) was rotated at the same speed as the gantry, creating a partial ring artifact where the angular extent of the artifact provides a direct indication of the temporal resolution. A demonstration of the effect of the evaluated 165 and 83 ms nominal temporal resolutions on motion artifacts in cardiac CT is provided using patient data.


Subject(s)
Algorithms , Artifacts , Radiographic Image Enhancement/methods , Radiographic Image Interpretation, Computer-Assisted/methods , Tomography, X-Ray Computed/methods , Humans , Phantoms, Imaging , Reproducibility of Results , Sensitivity and Specificity , Tomography, X-Ray Computed/instrumentation
12.
Invest Radiol ; 42(7): 491-8, 2007 Jul.
Article in English | MEDLINE | ID: mdl-17568271

ABSTRACT

PURPOSE: To investigate the influence of heart rate and temporal resolution on the assessment of global ventricular function with dual-source computed tomography (DSCT). MATERIALS AND METHODS: A dynamic cardiac phantom was repeatedly scanned with a DSCT scanner applying a standardized scan protocol at different heart rates, ranging from 40 to 140 bpm. Images were reconstructed with monosegmental and bisegmental algorithms using data from a single source and from both sources. Ventricular volumes and ejection fraction (EF) were computed by semiautomated analysis. Results were compared with the phantom's real volumes. Interscan, intraobserver, and interobserver variability were calculated. RESULTS: For single-source data reconstruction temporal resolution was fixed to 165 milliseconds, whereas dual-source image reconstructions resulted in a temporal resolution of 83 milliseconds (monosegmental) and 67.7+/-14.2 milliseconds (bisegmental), respectively. In general, deviation from the phantom's real volumes was less with dual-source data reconstruction when compared with single-source data reconstruction. Comparing dual-source data reconstruction with single-source data reconstruction, the percent deviation from the phantom's real volumes for EF was 0.7% (monosegmental), 0.7% (bisegmental), and 4.3% (single source), respectively. There was no correlation between heart rate and EF for dual-source data reconstruction (r=-0.168; r=-0.157), whereas a relevant correlation was observed for single-source data reconstruction (r=-0.844). Interscan, intraobserver, and interobserver variability for EF were 1.4%, 0.9%, and 0.3%, respectively. CONCLUSIONS: DSCT allows reliable quantification of global ventricular function independent of the heart rate. Multisegmental image reconstruction is not needed for DSCT assessment of global ventricular function.


Subject(s)
Heart/physiology , Image Processing, Computer-Assisted/instrumentation , Phantoms, Imaging , Stroke Volume , Tomography, X-Ray Computed/instrumentation , Ventricular Function , Algorithms , Heart Rate , Heart Ventricles/pathology , Humans , Pilot Projects
13.
Radiology ; 243(3): 775-84, 2007 Jun.
Article in English | MEDLINE | ID: mdl-17446525

ABSTRACT

PURPOSE: To prospectively compare the dose performance of a 64-channel multi-detector row computed tomographic (CT) scanner and a 64-channel dual-source CT scanner from the same manufacturer. MATERIALS AND METHODS: To minimize dose in the cardiac (dual-source) mode, the evaluated dual-source CT system uses a cardiac beam-shaping filter, three-dimensional adaptive noise reduction, heart rate-dependent pitch, and electrocardiographically based modulation of the tube current. Weighted CT dose index per 100 mAs was measured for the head, body, and cardiac beam-shaping filters. Kerma-length product was measured in the spiral cardiac mode at four pitch values and three electrocardiographic modulation temporal windows. Noise was measured in an anthropomorphic phantom. Data were compared with data from a 64-channel multi-detector row CT scanner. RESULTS: For the multi-detector row and dual-source CT systems, respectively, weighted CT dose index per 100 mAs was 14.2 and 12.2 mGy (head CT), 6.8 and 6.4 mGy (body CT), and 6.8 and 5.3 mGy (cardiac CT). In the spiral cardiac mode (no electrocardiographically based tube current modulation, 0.2 pitch), equivalent noise occurred at volume CT dose index values of 23.7 and 35.0 mGy (coronary artery calcium CT) and 58.9 and 61.2 mGy (coronary CT angiography) for multi-detector row CT and dual-source CT, respectively. The use of heart rate-dependent pitch values reduced volume CT dose index to 46.2 mGy (0.265 pitch), 34.0 mGy (0.36 pitch), and 26.6 mGy (0.46 pitch) compared with 61.2 mGy for 0.2 pitch. The use of electrocardiographically based tube current-modulation and temporal windows of 110, 210, and 310 msec further reduced volume CT dose index to 9.1-25.1 mGy, dependent on the heart rate. CONCLUSION: For electrocardiographically gated coronary CT angiography, image noise equivalent to that of multi-detector row CT can be achieved with dual-source CT at doses comparable to or up to a factor of two lower than the doses at multi-detector row CT, depending on heart rate of the patient.


Subject(s)
Equipment Failure Analysis , Tomography, X-Ray Computed/instrumentation , Whole-Body Counting/methods , Body Burden , Equipment Design , Humans , Radiation Dosage , Relative Biological Effectiveness , Risk Assessment , Risk Factors
14.
Eur Radiol ; 17(6): 1510-7, 2007 Jun.
Article in English | MEDLINE | ID: mdl-17151859

ABSTRACT

The aim of this study was to assess the feasibility of a differentiation of iodine from other materials and of different body tissues using dual energy CT. Ten patients were scanned on a SOMATOM Definition Dual Source CT (DSCT; Siemens, Forchheim, Germany) system in dual energy mode at tube voltages of 140 and 80 kVp and a ratio of 1:3 between tube currents. Weighted CT Dose Index ranged between 7 and 8 mGy, remaining markedly below reference dose values for the respective body regions. Image post-processing with three-material decomposition was applied to differentiate iodine or collagen from other tissue. The results showed that a differentiation and depiction of contrast material distribution is possible in the brain, the lung, the liver and the kidneys with or without the underlying tissue of the organ. In angiographies, bone structures can be removed from the dataset to ease the evaluation of the vessels. The differentiation of collagen makes it possible to depict tendons and ligaments. Dual energy CT offers a more specific tissue characterization in CT and can improve the assessment of vascular disease. Further studies are required to draw conclusions on the diagnostic value of the individual applications.


Subject(s)
Contrast Media/pharmacokinetics , Iohexol/analogs & derivatives , Radiographic Image Interpretation, Computer-Assisted/methods , Tomography, X-Ray Computed/methods , Algorithms , Diagnosis, Differential , Feasibility Studies , Female , Humans , Imaging, Three-Dimensional , Iohexol/pharmacokinetics , Male
15.
Invest Radiol ; 41(12): 898-903, 2006 Dec.
Article in English | MEDLINE | ID: mdl-17099429

ABSTRACT

OBJECTIVE: We sought to evaluate an electrocardiogram (ECG)-independent image reconstruction technique for coronary computed tomography (CT)-angiography based on cardiac motion. MATERIALS AND METHODS: The raw data from 20 patients was reconstructed with both an ECG-gated algorithm and a motion-dependent algorithm that calculates the cardiac motion-function directly from the CT raw data using a center of mass technique. Images were reconstructed in 5% steps over the R-R interval and the cardiac motion-cycle. For both approaches multiplanar reformations were created and the set of images with the least motion artifacts was used for the evaluation. Motion artifacts affecting the ascending aorta, the left main coronary artery and the entire course of the LAD, LCX and RCA were scored using a 5-point scale. RESULTS: The mean optimal reconstruction window was at 60% of the R-R interval and 30% of the cardiac motion cycle. A total of 73 of 100 vascular regions showed no motion artifacts in ECG-gated images, with the motion-synchronized algorithm only 41 regions were free of motion artifacts. The mean motion-score was 1.4 (+/-0.6) and 2.4 (+/-1.2) respectively (P < 0.05). CONCLUSION: In the currently implemented form the motion-gated algorithm is inferior to ECG-gated image reconstruction but can be used in patients with an incomplete or corrupt ECG-signal.


Subject(s)
Algorithms , Image Enhancement/methods , Image Interpretation, Computer-Assisted/methods , Motion , Myocardial Contraction , Tomography, X-Ray Computed , Artifacts , Electrocardiography , Heart Rate , Humans , Image Interpretation, Computer-Assisted/standards , Quality Control , Tomography, X-Ray Computed/methods
16.
Invest Radiol ; 41(5): 429-35, 2006 May.
Article in English | MEDLINE | ID: mdl-16625105

ABSTRACT

PURPOSE: We sought to investigate the influence of heart rate and temporal resolution on the assessment of left-ventricular (LV) function with multislice spiral computed tomography (CT). MATERIAL AND METHODS: A dynamic cardiac phantom was repeatedly scanned with a 64-slice CT scanner using a standardized scan protocol (64 x 0.6 mm, 120kV, 770mAs(eff), 330 milliseconds rotation time) at different simulated heart rates, ranging from 40 to 140 beats per minute. Images were reconstructed with an algorithm utilizing data from 1 to 4 cardiac cycles (RR intervals). Ejection fraction (EF), end-systolic, end-diastolic, and stroke volume as well as cardiac output were calculated. Results of the measurements were compared with the real volumes of the phantom. Interscan and intraobserver variability were calculated. RESULTS: Using a monosegmental reconstruction algorithm, the temporal resolution was fixed to 165 milliseconds. With bi-, tri-, and quad-segmental image reconstruction, mean temporal resolution was 128.3 +/- 33.2 milliseconds, 103.3 +/- 49.2 milliseconds, and 87.8 +/- 81.5 milliseconds, respectively. Multisegmental image reconstruction resulted in a lower deviation when comparing measured and real volumes. Using mono-, bi-, tri-, and quad-segmental image reconstruction, the percent deviation between measured and real values for EF was 8.2%, 4.5%, 3.3%, and 3.4%, respectively. Applying multisegmental image reconstruction with improved temporal resolution the deviation decreased with increasing heart rate when compared with mono-segmental image reconstruction. Interscan and intraobserver variability for EF were 1.1% and 1.9%, respectively. CONCLUSION: Enhanced temporal resolution improves the quantification of LV volumes in cardiac multislice spiral CT, enabling reliable assessment of LV volumes even at increased heart rates.


Subject(s)
Heart Rate/physiology , Phantoms, Imaging , Stroke Volume/physiology , Tomography, Spiral Computed/instrumentation , Ventricular Function, Left/physiology , Algorithms , Cardiac Volume/physiology , Humans , Radiographic Image Interpretation, Computer-Assisted
17.
Eur J Radiol ; 57(3): 331-5, 2006 Mar.
Article in English | MEDLINE | ID: mdl-16426789

ABSTRACT

UNLABELLED: Multi-detector computed tomography (CT) scanners, by virtue of their high temporal and spatial resolution, permit imaging of the coronary arteries. However, motion artifacts, especially in patients with higher heart rates, can impair image quality. We thus evaluated the performance of a new dual-source CT (DSCT) with a heart rate independent temporal resolution of 83 ms for the visualization of the coronary arteries in 14 consecutive patients. METHODS: Fourteen patients (mean age 61 years, mean heart rate 71 min(-1)) were studied by DSCT. The system combines two arrays of an X-ray tube plus detector (64 slices) mounted on a single gantry at an angle of 90 degrees With a rotation speed of 330 ms, a temporal resolution of 83 ms (one-quarter rotation) can be achieved independent of heart rate. For data acquisition, intraveous contrast agent was injected at a rate of 5 ml/s. Images were reconstructed with 0.75 slice thickness and 0.5 mm increment. The data sets were evaluated concerning visibility of the coronary arteries and occurrence of motion artifact. RESULTS: Visualization of the coronary arteries was successful in all patients. Most frequently, image reconstruction at 70% of the cardiac cycle provided for optimal image quality (50% of patients). Of a total of 226 coronary artery segments, 222 (98%) were visualized free of motion artifact. In summary, DSCT constitutes a promising new concept for cardiac CT. High and heart rate independent temporal resolution permits imaging of the coronary arteries without motion artifacts in a substantially increased number of patients as compared to earlier scanner generations. Larger and appropriately designed studies will need to determine the method's accuracy for detection of coronary artery stenoses.


Subject(s)
Coronary Angiography/methods , Coronary Disease/diagnostic imaging , Image Processing, Computer-Assisted , Tomography, X-Ray Computed/methods , Artifacts , Contrast Media , Electrocardiography , Heart Rate , Humans , Male , Middle Aged , Motion
18.
Eur Radiol ; 16(2): 256-68, 2006 Feb.
Article in English | MEDLINE | ID: mdl-16341833

ABSTRACT

We present a performance evaluation of a recently introduced dual-source computed tomography (DSCT) system equipped with two X-ray tubes and two corresponding detectors, mounted onto the rotating gantry with an angular offset of 90 degrees . We introduce the system concept and derive its consequences and potential benefits for electrocardiograph [corrected] (ECG)-controlled cardiac CT and for general radiology applications. We evaluate both temporal and spatial resolution by means of phantom scans. We present first patient scans to illustrate the performance of DSCT for ECG-gated cardiac imaging, and we demonstrate first results using a dual-energy acquisition mode. Using ECG-gated single-segment reconstruction, the DSCT system provides 83 ms temporal resolution independent of the patient's heart rate for coronary CT angiography (CTA) and evaluation of basic functional parameters. With dual-segment reconstruction, the mean temporal resolution is 60 ms (minimum temporal resolution 42 ms) for advanced functional evaluation. The z-flying focal spot technique implemented in the evaluated DSCT system allows 0.4 mm cylinders to be resolved at all heart rates. First clinical experience shows a considerably increased robustness for the imaging of patients with high heart rates. As a potential application of the dual-energy acquisition mode, the automatic separation of bones and iodine-filled vessels is demonstrated.


Subject(s)
Coronary Angiography/instrumentation , Coronary Stenosis/diagnosis , Echocardiography, Four-Dimensional/instrumentation , Electrocardiography/instrumentation , Image Enhancement/instrumentation , Image Processing, Computer-Assisted/instrumentation , Technology Assessment, Biomedical , Tomography, Spiral Computed/instrumentation , Diastole/physiology , Equipment Design , Heart Rate/physiology , Humans , Male , Middle Aged , Phantoms, Imaging , Sensitivity and Specificity , Systole/physiology
19.
Radiology ; 235(3): 756-73, 2005 Jun.
Article in English | MEDLINE | ID: mdl-15833981

ABSTRACT

The introduction in 1998 of multi-detector row computed tomography (CT) by the major CT vendors was a milestone with regard to increased scan speed, improved z-axis spatial resolution, and better utilization of the available x-ray power. In this review, the general technical principles of multi-detector row CT are reviewed as they apply to the established four- and eight-section systems, the most recent 16-section scanners, and future generations of multi-detector row CT systems. Clinical examples are used to demonstrate both the potential and the limitations of the different scanner types. When necessary, standard single-section CT is referred to as a common basis and starting point for further developments. Another focus is the increasingly important topic of patient radiation exposure, successful dose management, and strategies for dose reduction. Finally, the evolutionary steps from traditional single-section spiral image-reconstruction algorithms to the most recent approaches toward multisection spiral reconstruction are traced.


Subject(s)
Image Processing, Computer-Assisted , Tomography, X-Ray Computed/methods , Electrocardiography , Equipment Design , Forecasting , Humans , Image Processing, Computer-Assisted/trends , Mathematics , Radiation Dosage , Tomography, X-Ray Computed/instrumentation , Tomography, X-Ray Computed/trends
20.
Phys Med Biol ; 49(11): 2209-18, 2004 Jun 07.
Article in English | MEDLINE | ID: mdl-15248573

ABSTRACT

A new 3D reconstruction scheme, weighted filtered backprojection (WFBP) for multirow spiral CT based on an extension of the two-dimensional SMPR algorithm is described and results are presented. In contrast to other 3D algorithms available, the algorithm makes use of all available data for all pitch values. The algorithm is a FBP algorithm: linear convolution of the parallel data along the row direction followed by a 3D backprojection. Data usage for arbitrary pitch values is maintained through a weighting scheme which takes into account redundant data. If proper row weighting is applied, the image quality is superior to the image quality of the SMPR algorithm.


Subject(s)
Algorithms , Imaging, Three-Dimensional/methods , Information Storage and Retrieval/methods , Radiographic Image Interpretation, Computer-Assisted/methods , Tomography, Spiral Computed/methods , Numerical Analysis, Computer-Assisted , Radiographic Image Enhancement/methods , Reproducibility of Results , Sensitivity and Specificity , Signal Processing, Computer-Assisted
SELECTION OF CITATIONS
SEARCH DETAIL
...