Omental wrapping is a common cause of peritoneal catheter malfunction. This diagnosis should be confirmed by radiography before proceeding with surgical omentectomy. We report two cases of peritoneal dialysis (PD) catheter outflow obstruction from our developing PD program in Grenada, in which contrast studies accurately diagnosed omental wrap, allowing for prompt surgical correction. In both cases, the contrast study indicated the presence of omental wrapping, confirmed at time of surgical correction. Radiographic features of omental wrap are distinctive, which allows for reliable differentiation from other causes of obstruction. Radiographic contrast study reliably diagnoses the cause of peritoneal catheter obstruction, permitting prompt diagnosis and treatment. This is vital for regions with limited access to haemodialysis.