1.
J Mech Behav Biomed Mater ; 153: 106507, 2024 May.
Article in English | MEDLINE | ID: mdl-38503082

ABSTRACT

Polyolefins exhibit robust mechanical and chemical properties and can be applied in the medical field, e.g. for the manufacturing of dentures. Despite their wide range of applications, they are rarely used in extrusion-based printing due to their warpage tendency. The aim of this study was to investigate and reduce the warpage of polyolefins compared to commonly used filaments after additive manufacturing (AM) and sterilization using finite element simulation. Three types of filaments were investigated: a medical-grade polypropylene (PP), a glass-fiber reinforced polypropylene (PP-GF), and a biocopolyester (BE) filament, and they were compared to an acrylic resin (AR) for material jetting. Square specimens, standardized samples prone to warpage, and denture bases (n = 10 of each group), as clinically relevant and anatomically shaped reference, were digitized after AM and steam sterilization (134 °C). To determine warpage, the volume underneath the square specimens was calculated, while the deviations of the denture bases from the printing file were measured using root mean square (RMS) values. To reduce the warpage of the PP denture base, a simulation of the printing file based on thermomechanical calculations was performed. Statistical analysis was conducted using the Kruskal-Wallis test, followed by Dunn's test for multiple comparisons. The results showed that PP exhibited the greatest warpage of the square specimens after AM, while PP-GF, BE, and AR showed minimal warpage before sterilization. However, warpage increased for PP-GF, BE and AR during sterilization, whereas PP remained more stable. After AM, denture bases made of PP showed the highest warpage. Through simulation-based optimization, warpage of the PP denture base was successfully reduced by 25%. In contrast to the reference materials, PP demonstrated greater dimensional stability during sterilization, making it a potential alternative for medical applications. Nevertheless, reducing warpage during the cooling process after AM remains necessary, and simulation-based optimization holds promise in addressing this issue.


Subject(s)
Polypropylenes , Steam , Polyenes , Acrylic Resins/chemistry , Sterilization
2.
J Mech Behav Biomed Mater ; 152: 106418, 2024 Apr.
Article in English | MEDLINE | ID: mdl-38295512

ABSTRACT

Different printing technologies can be used for prosthetically oriented implant placement, however the influence of different printing orientations and steam sterilization remains unclear. In particular, no data is available for the novel technology Continuous Liquid Interface Production. The objective was to evaluate the dimensional accuracy of surgical guides manufactured with different printing techniques in vertical and horizontal printing orientation before and after steam sterilization. A total of 80 surgical guides were manufactured by means of continuous liquid interface production (CLIP; material: Keyguide, Keyprint), digital light processing (DLP; material: Luxaprint Ortho, DMG), stereolithography (SLA; Surgical guide, Formlabs), and fused filament fabrication (FFF; material: Clear Base Support, Arfona) in vertical and horizontal printing orientation (n = 10 per subgroup). Spheres were included in the design to determine the coordinates of 17 reference points. Each specimen was digitized with a laboratory scanner after additive manufacturing (AM) and after steam sterilization (134 °C). To determine the accuracy, root mean square values (RMS) were calculated and coordinates of the reference points were recorded. Based on the measured coordinates, deviations of the reference points and relevant distances were calculated. Paired t-tests and one-way ANOVA were applied for statistical analysis (significance p < 0.05). After AM, all printing technologies showed comparable high accuracy, with an increased deviation in z-axis when printed horizontally. After sterilization, FFF printed surgical guides showed distinct warpage. The other subgroups showed no significant differences regarding the RMS of the corpus after steam sterilization (p > 0.05). Regarding reference points and distances, CLIP showed larger deviations compared to SLA in both printing orientations after steam sterilization, while DLP manufactured guides were the most dimensionally stable. In conclusion, the different printing technologies and orientations had little effect on the manufacturing accuracy of the surgical guides before sterilization. However, after sterilization, FFF surgical guides exhibited significant deformation making their clinical use impossible. CLIP showed larger deformations due to steam sterilization than the other photopolymerizing techniques, however, discrepancies may be considered within the range of clinical acceptance. The influence on the implant position remains to be evaluated.


Subject(s)
Steam , Stereolithography , Analysis of Variance , Cytoskeleton , Sterilization
3.
Clin Oral Implants Res ; 35(4): 377-385, 2024 Apr.
Article in English | MEDLINE | ID: mdl-38170349

ABSTRACT

OBJECTIVES: The aim of this study was to evaluate the effect of sterilization on the retention forces of lithium disilicate (LD) and polymer-infiltrated ceramic network (PICN) crowns bonded to titanium base (Ti-base) abutments. MATERIALS AND METHODS: Forty LD and 40 PICN crowns were milled and then bonded to 80 Ti-bases with two resin composite cements: Multilink Hybrid Abutment (mh) and Panavia V5 (pv) for a total of 8 groups (n = 10). Half of the specimens (test) underwent an autoclaving protocol (pressure 1.1 bar, 121°C, 20.5 min) and the other half not (control). Restorations were screw-retained to implants, and retention forces (N) were measured with a pull-off testing machine. The surfaces of the Ti-bases and the crowns were inspected for the analysis of the integrity of the marginal bonding interface and failure mode. Student's t-test, chi-square test, and univariate linear regression model were performed to analyze the data (α = 0.05). RESULTS: The mean pull-off retention forces ranged from 487.7 ± 73.4 N to 742.2 ± 150.3 N. Sterilized groups showed statistically significant overall higher maximum retention forces (p < .05), except for one combination (LD + mh). Sterilization led to an increased presence of marginal gaps and deformities compared to no-sterilization (p < .001), while no statistically significant relationship was found between failure mode and sterilization (p > .05). CONCLUSIONS: Sterilization may have a beneficial effect on the retention forces of LD and PICN crowns bonded to titanium base abutments, although it may negatively influence the integrity of the marginal bonding interface.


Subject(s)
Polymers , Titanium , Dental Porcelain , Crowns , Materials Testing , Zirconium , Ceramics , Dental Stress Analysis , Dental Abutments , Computer-Aided Design
4.
Int J Comput Assist Radiol Surg ; 19(3): 591-599, 2024 Mar.
Article in English | MEDLINE | ID: mdl-37523011

ABSTRACT

PURPOSE: The aim of the study was to assess the deviation between clinical implant axes (CIA) determined by a surgeon during preoperative planning and reconstructed tooth axes (RTA) of missing teeth which were automatically computed by a previously introduced anatomical SSM. METHODS: For this purpose all available planning datasets of single-implant cases of our clinic, which were planned with coDiagnostix Version 9.9 between 2018 and 2021, were collected for retrospective investigation. Informed consent was obtained. First, the intraoral scans of implant patients were annotated and subsequently analyzed using the SSM. The RTA, computed by the SSM, was then projected into the preoperative planning dataset. The amount and direction of spatial deviation between RTA and CIA were then measured. RESULTS: Thirty-five patients were implemented. The mean distance between the occlusal entry point of anterior and posterior implants and the RTA was 0.99 mm ± 0.78 mm and 1.19 mm ± 0.55, respectively. The mean angular deviation between the CIA of anterior and posterior implants and the RTA was 12.4° ± 3.85° and 5.27° ± 2.97° respectively. The deviations in anterior implant cases were systematic and could be corrected by computing a modified RTA (mRTA) with decreased deviations (0.99 mm ± 0.84 and 4.62° ± 1.95°). The safety distances of implants set along the (m)RTA to neighboring teeth were maintained in 30 of 35 cases. CONCLUSION: The RTA estimated by the SSM revealed to be a viable implant axis for most of the posterior implant cases. As there are natural differences between the anatomical tooth axis and a desirable implant axis, modifications were necessary to correct the deviations which occurred in anterior implant cases. However, the presented approach is not applicable for clinical use and always requires manual optimization by the planning surgeon.


Subject(s)
Dental Implants , Surgery, Computer-Assisted , Tooth , Humans , Cone-Beam Computed Tomography , Retrospective Studies , Prostheses and Implants , Computer-Aided Design , Imaging, Three-Dimensional
5.
Clin Oral Implants Res ; 34 Suppl 26: 86-103, 2023 Sep.
Article in English | MEDLINE | ID: mdl-37750526

ABSTRACT

OBJECTIVE: The objectives of the study were to assess the survival, failure, and technical complication rates of implant-supported fixed dental prosthesis (iFDPs) with pontic or splinted crown (iSp C) designs in the posterior area and compare the influence of prosthetic materials and prosthetic design on the outcomes. METHODS: Electronic and manual searches were performed to identify randomized-, prospective-, and retrospective clinical trials with follow-up time of ≥12 months, evaluating the clinical outcomes of posterior iFDPs with pontic or iSp Cs. Survival and complication rates were analyzed using robust Poisson's regression models. RESULTS: Thirty-two studies reporting on 42 study arms were included in the present systematic review. The meta-analysis of the included studies indicated estimated 3-year survival rates of 98.3% (95%CI: 95.6-99.3%) for porcelain-fused-to-metal (PFM) iFDPs, 97.5% (95%CI: 95.5-98.7%) for veneered zirconia (Zr) iFDPs with pontic, 98.9% (95%CI: 96.8-99.6%) for monolithic or micro-veneered zirconia iFDPs with pontic, and 97.0% (95%CI: 84.8-99.9%) for lithium disilicate iFDPs with pontics. The survival rates for different material combination showed no statistically significant differences. Veneered restorations, overall, showed significantly (p < .01) higher ceramic fracture and chipping rates compared with monolithic restorations. Furthermore, there was no significant difference in survival rates (98.3% [95%CI: 95.6-99.3%] vs. 99.1% [95%CI: 97.6-99.7%]) and overall complication rates between PFM iFDPs with pontic and PFM iSp Cs. CONCLUSIONS: Based on the data identified by this systematic review, PFM, veneered Zr, and monolithic Zr iFDPs with pontic and iSp Cs showed similarly high short-term survival rates in the posterior area. Veneered restorations exhibit ceramic chipping more often than monolithic restorations, with the highest fracture rate reported for veneered Zr iFDPs.


Subject(s)
Dental Prosthesis Design , Dental Restoration Failure , Prospective Studies , Retrospective Studies , Dental Porcelain , Ceramics , Zirconium , Crowns , Dental Prosthesis, Implant-Supported
6.
IEEE Trans Pattern Anal Mach Intell ; 45(9): 10745-10759, 2023 09.
Article in English | MEDLINE | ID: mdl-37015129

ABSTRACT

Recent advances in transformer-based architectures have shown promise in several machine learning tasks. In the audio domain, such architectures have been successfully utilised in the field of speech emotion recognition (SER). However, existing works have not evaluated the influence of model size and pre-training data on downstream performance, and have shown limited attention to generalisation, robustness, fairness, and efficiency. The present contribution conducts a thorough analysis of these aspects on several pre-trained variants of wav2vec 2.0 and HuBERT that we fine-tuned on the dimensions arousal, dominance, and valence of MSP-Podcast, while additionally using IEMOCAP and MOSI to test cross-corpus generalisation. To the best of our knowledge, we obtain the top performance for valence prediction without use of explicit linguistic information, with a concordance correlation coefficient (CCC) of. 638 on MSP-Podcast. Our investigations reveal that transformer-based architectures are more robust compared to a CNN-based baseline and fair with respect to gender groups, but not towards individual speakers. Finally, we show that their success on valence is based on implicit linguistic information, which explains why they perform on-par with recent multimodal approaches that explicitly utilise textual information. To make our findings reproducible, we release the best performing model to the community.


Subject(s)
Algorithms , Speech , Emotions , Machine Learning
7.
J Dent ; 132: 104482, 2023 05.
Article in English | MEDLINE | ID: mdl-36931618

ABSTRACT

OBJECTIVES: This in vitro study aimed to assess the influence of the planning software and design of the surgical template on both trueness and precision of static computer assisted implant surgery (sCAIS) performed using guides fabricated using material extrusion (ME). METHODS: Three-dimensional radiographic and surface scans of a typodont were aligned using two planning software (coDiagnostiX, CDX; ImplantStudio, IST) to virtually position the two adjacent oral implants. Thereafter, surgical guides were fabricated with either an original (O) or modified (M) design with reduced occlusal support and were sterilized. Forty surgical guides were used to install 80 implants equally distributed amongst four groups: CDX-O, CDX-M, IST-O, and IST-M. Thereafter, the scan bodies were adapted to the implants and digitized. Finally, inspection software was used to assess discrepancies between the planned and final positions at the implant shoulder and main axis level. Multilevel mixed-effects generalised linear models were used for statistical analyses (p = 0.05). RESULTS: In terms of trueness, the largest average vertical deviations (0.29 ± 0.07 mm) were be assessed for CDX-M. Overall, vertical errors were dependant on the design (O < M; p ≤ 0.001). Furthermore, in horizontal direction, the largest mean discrepancy was 0.32 ± 0.09 mm (IST-O) and 0.31 ± 0.13 mm (CDX-M). CDX-O was superior compared to IST-O (p = 0.003) regarding horizontal trueness. The average deviations regarding the main implant axis ranged between 1.36 ± 0.41° (CDX-O) and 2.63 ± 0.87° (CDX-M). In terms of precision, mean standard deviation intervals of ≤ 0.12 mm (IST-O and -M) and ≤ 1.09° (CDX-M) were calculated. CONCLUSIONS: Implant installation with clinically acceptable deviations is possible with ME surgical guides. Both evaluated variables affected trueness and precision with negligible differences. CLINICAL SIGNIFICANCE: The planning system and design influenced the accuracy of implant installation using ME-based surgical guides. Nevertheless, discrepancies were ≤ 0.32 mm and ≤ 2.63°, which may be considered within the range of clinical acceptance. ME should be further investigated as an alternative to the more expensive and time-consuming 3D printing technologies.


Subject(s)
Dental Implants , Dental Implantation, Endosseous , Computer-Aided Design , Imaging, Three-Dimensional , Cone-Beam Computed Tomography , Software , Computers
8.
J Funct Biomater ; 14(3)2023 Feb 24.
Article in English | MEDLINE | ID: mdl-36976047

ABSTRACT

The aim of this in vitro study was to evaluate the long-term stability of one-piece diameter reduced zirconia oral implants under the influence of loading and artificial aging in a chewing simulator as well as the fracture load in a static loading test. Thirty-two one-piece zirconia implants with a diameter of 3.6 mm were embedded according to the ISO 14801:2016 standard. The implants were divided into four groups of eight implants. The implants of group DLHT were dynamically loaded (DL) in a chewing simulator for 107 cycles with a load of 98 N and simultaneously hydrothermally aged (HT) using a hot water bath at 85 °C. Group DL was only subjected to dynamic loading and group HT was exclusively subjected to hydrothermal aging. Group 0 acted as a control group: no dynamical loading, no hydrothermal ageing. After exposure to the chewing simulator, the implants were statically loaded to fracture in a universal testing machine. To evaluate group differences in the fracture load and bending moments, a one-way ANOVA with Bonferroni correction for multiple testing was performed. The level of significance was set to p < 0.05. In the static loading test, group DLHT showed a mean fracture load of 511 N, group DL of 569 N, group HT of 588 N and control group 0 of 516 N. The average bending moments had the following values: DLHT: 283.5 Ncm; DL: 313.7 Ncm; HT: 324.4 Ncm; 0: 284.5 Ncm. No significant differences could be found between the groups. Hydrothermal aging and/or dynamic loading had no significant effect on the stability of the one-piece diameter reduced zirconia implants (p > 0.05). Within the limits of this investigation, it can be concluded that dynamic loading, hydrothermal aging and the combination of loading and aging did not negatively influence the fracture load of the implant system. The artificial chewing results and the fracture load values indicate that the investigated implant system seems to be able to resist physiological chewing forces also over a long service period.

9.
J Funct Biomater ; 14(2)2023 Feb 19.
Article in English | MEDLINE | ID: mdl-36826915

ABSTRACT

The intention of this 5-year prospective cohort investigation was to clinically and radiographically investigate the outcomes of a one-piece zirconia implant system for single tooth replacement. Sixty-five patients received a total of 66 single-tooth implants. All implants immediately received temporary restorations and were finally restored with all-ceramic crowns. Follow-ups were performed at the prosthetic delivery, after 1, 3, and 5 years. Peri-implant and dental soft-tissue parameters were evaluated and patient-reported outcomes recorded. To monitor peri-implant bone remodelling, standardised radiographs were taken at the implant insertion and at the 1-, 3-, and 5-year follow-ups. In the course of 5 years, 14 implants were lost, resulting in a cumulative implant survival rate of 78.2%. The mean marginal bone loss from the implant insertion to the 5-year follow-up amounted to 1.12 mm. Probing depth, clinical attachment level, bleeding, and plaque index increased over time. In 91.5% of the implants, the papilla index showed levels of 1 or 2, respectively. At the end of the study, the patient satisfaction was higher compared to the pre-treatment measurements. Due to the low survival rate after five years and the noticeably high frequency of advanced bone loss observed in this study, the implant has not met the launch criteria, as it would have not been recommended for routine clinical use.

10.
J Funct Biomater ; 14(1)2023 Jan 13.
Article in English | MEDLINE | ID: mdl-36662092

ABSTRACT

The objective was to investigate the clinical and radiological outcome of one-piece zirconia oral implants to support three-unit fixed dental prostheses (FDP) after three years in function. Twenty-seven patients were treated with a total of 54 implants in a one-stage surgery and immediate provisionalization. Standardized radiographs were taken at implant placement, after one year and after three years, to evaluate peri-implant bone loss. Soft-tissue parameters were also assessed. Linear mixed regression models as well as Wilcoxon Signed Rank tests were used for analyzing differences between groups and time points (p < 0.05). At the three-year evaluation, one implant was lost, resulting in a cumulative survival rate of 98.1%. The mean marginal bone loss amounted to 2.16 mm. An implant success grade I of 52% (bone loss of ≤2 mm) and success grade II of 61% (bone loss of ≤3 mm) were achieved. None of the evaluated baseline parameters affected bone loss. The survival rate of the zirconia implants was comparable to market-available titanium implants. However, an increased marginal bone loss was observed with a high peri-implantitis incidence and a resulting low implant success rate.

11.
Clin Oral Implants Res ; 34(2): 105-115, 2023 Feb.
Article in English | MEDLINE | ID: mdl-36444693

ABSTRACT

OBJECTIVE: To investigate the fracture strength and potential phase transformation of an injection-molded two-piece zirconia implant restored with a zirconia abutment after loading and/or aging. METHODS: Thirty-two two-piece zirconia implants (4.0 mm diameter) restored with zirconia abutments were embedded according to ISO 14801 and divided into four groups (n = 8/group): Three groups were either exclusively hydrothermally treated (group HT; 85°C), dynamically loaded (group DL; 107 cycles; 98 N), or subjected to both treatments simultaneously (group DL/HT). One group remained untreated (group 0). A sample from each group was cross-sectioned and examined by scanning electron microscopy for possible crystal phase transformation. The remaining samples were then loaded to fracture in a static loading test. A one-way ANOVA was used for statistical analyses. RESULTS: During dynamic loading, three implants of group DL and six implants of group DL/HT fractured at a load of 98 N. The fracture strength of group DL/HT (108 ± 141 Ncm) was significantly reduced compared to the other groups (group 0: 342 ± 36 Ncm; HT: 363 ± 49 Ncm; DL: 264 ± 198 Ncm) (p < .05). Fractures from group 0 and HT occurred at both implant and abutment level, whereas implants from group DL and DL/HT fractured only at implant level. A shallow monoclinic transformation zone of approximately 2 µm was observed following hydrothermal treatment. CONCLUSIONS: Within the limitations of this study, it can be concluded that dynamic loading and the combination of loading and aging reduced the fracture strength of the implant abutment combination. Hydrothermal treatment caused a shallow transformation zone which had no influence on the fracture strength.


Subject(s)
Dental Implants , Flexural Strength , Dental Implant-Abutment Design , Materials Testing , Titanium/chemistry , Zirconium/chemistry , Dental Stress Analysis , Dental Abutments , Dental Restoration Failure
12.
Int J Prosthodont ; 36(5): 651, 2023 Nov 01.
Article in English | MEDLINE | ID: mdl-36288489

ABSTRACT

PURPOSE: To assess the influence of the bonding system and restorative material on the marginal integrity and pull-off forces of monolithic all-ceramic crowns bonded to titanium base (ti-base) abutments. MATERIALS AND METHODS: A total of 108 ti-bases were sandblasted and divided into nine experimental groups (n = 12) according to the combination of crown material (polymer-infiltrated ceramic-network [PI], lithium-disilicate [LD], and zirconia [ZI]) and bonding system (Multilink Hybrid-Abutment [MH], Panavia V5 [PV], RelyX Ul5mate [RU]) with the respective primers. After bonding the crowns to the ti-base abutments, the restorations were screw-retained on implants and thermomechanically aged (1,200,000 cycles, 49 N, 1.67 Hz, 5 to 55°C). Marginal integrity and bonding failures were evaluated under a light microscope, and pull-off forces (N) were calculated. Chi-square tests for marginal integrity as well as one-way and two-way ANOVA statistical tests for pull-off forces were applied (a = .05). RESULTS: PI presented higher marginal integrity than LD (P = .023). Bonding system PV revealed higher marginal integrity than MH (P =.005) and RU (P =.029). Differences in pull-off forces were found between restorative material and resin cements (P < .001), with the highest values for ZI + RU (598 ± 192 N), PI + PV (545 ± 114 N), LD + MH (532 ± 116 N), and PI + RU (528 ± 81 N). Specimens with marginal integrity revealed higher pull-off forces than those with alteration (P = .006). Specimens presenting bonding failures (micromovements) showed lower pull-off forces than those without bonding failures (P < .001). CONCLUSIONS: The tested CAD/CAM materials show favorable bonding performances with different bonding systems, nevertheless for each restorative material a specific bonding system has to be recommended. Int J Prosthodont 2023;36:e88-e102.


Subject(s)
Dental Materials , Titanium , Crowns , Ceramics , Resin Cements , Materials Testing , Zirconium , Computer-Aided Design , Dental Stress Analysis , Dental Abutments
13.
Int J Comput Assist Radiol Surg ; 17(10): 1957-1968, 2022 Oct.
Article in English | MEDLINE | ID: mdl-35902422

ABSTRACT

PURPOSE: Modern virtual implant planning is a time-consuming procedure, requiring a careful assessment of prosthetic and anatomical factors within a three-dimensional dataset. In order to facilitate the planning process and provide additional information, this study examines a statistical shape model (SSM) to compute the course of dental roots based on a surface scan. MATERIAL AND METHODS: Plaster models of orthognathic patients were scanned and superimposed with three-dimensional data of a cone-beam computer tomography (CBCT). Based on the open-source software "R", including the packages Morpho, mesheR, Rvcg and RvtkStatismo, an SSM was generated to estimate the tooth axes. The accuracy of the calculated tooth axes was determined using a leave-one-out cross-validation. The deviation of tooth axis prediction in terms of angle or horizontal shift is described with mean and standard deviation. The planning dataset of an implant surgery patient was additionally analyzed using the SSM. RESULTS: 71 datasets were included in this study. The mean angle between the estimated tooth-axis and the actual tooth-axis was 7.5 ± 4.3° in the upper jaw and 6.7 ± 3.8° in the lower jaw. The horizontal deviation between the tooth axis and estimated axis was 1.3 ± 0.8 mm close to the cementoenamel junction, and 0.7 ± 0.5 mm in the apical third of the root. Results for models with one missing tooth did not differ significantly. In the clinical dataset, the SSM could give a reasonable aid for implant positioning. CONCLUSIONS: With the presented SSM, the approximate course of dental roots can be predicted based on a surface scan. There was no difference in predicting the tooth axis of existent or missing teeth. In clinical context, the estimation of tooth axes of missing teeth could serve as a reference for implant positioning. However, a higher number of training data must be achieved to obtain increasing accuracy.


Subject(s)
Dental Implants , Surgery, Computer-Assisted , Computer-Aided Design , Cone-Beam Computed Tomography/methods , Feasibility Studies , Humans , Imaging, Three-Dimensional , Mandible , Maxilla , Surgery, Computer-Assisted/methods
14.
J Clin Med ; 11(9)2022 Apr 24.
Article in English | MEDLINE | ID: mdl-35566508

ABSTRACT

Adequate soft-tissue dimensions have been shown to be crucial for the long-term success of dental implants. To date, there is evidence that placement of dental implants should only be conducted in an area covered with attached gingiva. Modern implant planning software does not visualize soft-tissue dimensions. This study aims to calculate the course of the mucogingival borderline (MG-BL) using statistical shape models (SSM). Visualization of the MG-BL allows the practitioner to consider the soft tissue supply during implant planning. To deploy an SSM of the MG-BL, healthy individuals were examined and the intra-oral anatomy was captured using an intra-oral scanner (IOS). The empirical anatomical data was superimposed and analyzed by principal component analysis. Using a Leave-One-Out Cross Validation (LOOCV), the prediction of the SSM was compared with the original anatomy extracted from IOS. The median error for MG-BL reconstruction was 1.06 mm (0.49-2.15 mm) and 0.81 mm (0.38-1.54 mm) for the maxilla and mandible, respectively. While this method forgoes any technical work or additional patient examination, it represents an effective and digital method for the depiction of soft-tissue dimensions. To achieve clinical applicability, a higher number of datasets has to be implemented in the SSM.

15.
Sci Rep ; 12(1): 7391, 2022 05 05.
Article in English | MEDLINE | ID: mdl-35513701

ABSTRACT

Extrusion-based printing enables simplified and economic manufacturing of surgical guides for oral implant placement. Therefore, the cytotoxicity of a biocopolyester (BE) and a polypropylene (PP), intended for the fused filament fabrication of surgical guides was evaluated. For comparison, a medically certified resin based on methacrylic esters (ME) was printed by stereolithography (n = 18 each group). Human gingival keratinocytes (HGK) were exposed to eluates of the tested materials and an impedance measurement and a tetrazolium assay (MTT) were performed. Modulations in gene expression were analyzed by quantitative PCR. One-way ANOVA with post-hoc Tukey tests were applied. None of the materials exceeded the threshold for cytotoxicity (< 70% viability in MTT) according to ISO 10993-5:2009. The impedance-based cell indices for PP and BE, reflecting cell proliferation, showed little deviations from the control, while ME caused a reduction of up to 45% after 72 h. PCR analysis after 72 h revealed only marginal modulations caused by BE while PP induced a down-regulation of genes encoding for inflammation and apoptosis (p < 0.05). In contrast, the 72 h ME eluate caused an up-regulation of these genes (p < 0.01). All evaluated materials can be considered biocompatible in vitro for short-term application. However, long-term contact to ME might induce (pro-)apoptotic/(pro-)inflammatory responses in HGK.


Subject(s)
Polymers , Stereolithography , Gingiva , Humans , Keratinocytes , Polypropylenes/toxicity
16.
Int J Comput Dent ; 25(4): 349-359, 2022 Nov 25.
Article in English | MEDLINE | ID: mdl-35072424

ABSTRACT

PURPOSE: Creating wax-ups of missing teeth for backward planning in implant surgery is a complex and time-consuming process. To facilitate implant-planning procedures, the automatic generation of a virtual wax-up would be useful. In the present study, the reconstruction of missing teeth in partially edentulous patients was performed automatically using newly developed software. The accuracy was investigated in order to test its clinical applicability. MATERIALS AND METHODS: This study presents a new method for creating an automatic virtual wax-up, which could serve as a basic tool in modern implant-planning procedures. First, a statistical shape model (SSM) based on 76 maxillary and mandibular arch scans from dentally healthy individuals was generated. Then, artificially generated tooth gaps were reconstructed. The accuracy of the workflow was evaluated on a separate testing sample of 10 individuals with artificially created tooth gaps given as a median deviation, in millimeters. Scans of three clinical cases with partial edentulism were equally reconstructed using the SSM and compared with the final prosthodontic work. RESULTS: The reconstruction of the artificial tooth gaps could be performed with the following median reconstruction accuracy: gap 21 with 0.15 mm; gap 27 with 0.20 mm; gap 34 with 0.22 mm: gap 36 with 0.22 mm; gaps 12 to 22 with 0.22 mm; gaps 34 to 36 with 0.22 mm. A scenario for an almost edentulous mandible with all teeth missing except teeth 33 and 43 could be reconstructed with a median reconstruction accuracy of 0.37 mm. The median tooth gap deviation of the SSM-based reconstruction in clinical cases differed from the final inserted prosthodontic teeth by 0.49 to 0.86 mm in median. CONCLUSION: A first feasibility of creating virtual wax-ups using an SSM could be shown. Artificially generated tooth gaps could be reconstructed close to the original with the proposed workflow. In the clinical cases, the SSM proposes an anatomical reconstruction, which does not yet consider prosthodontic aspects. To obtain clinical use, contact with antagonist teeth must be considered and more training data must be implemented. However, the presented method offers a fast and viable way for the approximate placement of missing crowns. This could be used in a digital planning workflow when implant position must be determined. (Int J Comput Dent 2022;25(4):349-0; doi: 10.3290/j.ijcd.b2599407).


Subject(s)
Dental Implants , Mouth, Edentulous , Humans , Prosthodontics , Models, Statistical , Crowns
17.
Clin Oral Implants Res ; 32(11): 1288-1298, 2021 Nov.
Article in English | MEDLINE | ID: mdl-34352139

ABSTRACT

OBJECTIVE: To evaluate the influence of artificial aging on the transformation propagation and fracture resistance of zirconia implants. METHODS: One-piece (with integrated implant abutment, 1P; regular diameter [4.1mm]; n = 16) and two-piece (with separate implant abutment, 2P; wide diameter [5 mm]; n = 16) zirconia implants were embedded according to ISO 14801. A two-piece titanium-zirconium implant (Ti-Zr; 4.1 mm diameter) served as a control (n = 16). One subgroup (n = 8) of each system was simultaneously dynamically loaded (107 cycles; 98N) and hydrothermally aged (85°C, 58 days), while the other subgroup (n = 8) remained untreated. Finally, specimens were statically loaded to fracture. Potential crystal phase transformation was examined at cross sections using scanning electron microscopy (SEM). A multivariate linear regression model was applied for statistical analyses. RESULTS: The fracture resistance of 1P (1,117 [SD = 38] N; loaded/aged: 1,009 [60] N), 2P (850 [36] N; loaded/aged: 799 [84] N), and Ti-Zr implants (1,338 [205] N; loaded/aged: 1,319 [247] N) was not affected significantly by loading/aging (p = .171). However, when comparing the systems, they revealed significant differences independent of loading/aging (p ≤ .001). Regarding the crystal structure, a transformation zone was observed in SEM images of 1P only after aging, while 2P showed a transformation zone even before aging. After hydrothermal treatment, an increase of this monoclinic layer was observed in both systems. CONCLUSIONS: The Ti-Zr control implant showed higher fracture resistance compared to both zirconia implants. Loading/aging had no significant impact on the fracture resistance of both zirconia implants. The wide-body 2P zirconia implant was weaker than the regular body 1P implant.


Subject(s)
Dental Implants , Zirconium , Dental Abutments , Dental Implant-Abutment Design , Dental Restoration Failure , Dental Stress Analysis , Materials Testing , Titanium
18.
Clin Oral Implants Res ; 32(11): 1263-1273, 2021 Nov.
Article in English | MEDLINE | ID: mdl-34448240

ABSTRACT

OBJECTIVES: Soft-tissue volume augmentation treatments do not provide the satisfactory long-term functional and esthetic outcomes. The aim of the study was to develop a standardized digital procedure to design individual soft-tissue substitutes (STS) and apply mathematical modeling to obtain average shape STS for single posterior tooth defects. MATERIAL AND METHODS: Thirty-three casts from 30 patients were scanned. STS were designed with a computer-aided design software and a systematic procedure standardized the measurements across all STS using 3D-analysis software. The occlusal, mesial-distal, and buccal-lingual planes were defined to partition, each STS and produce a mesh. The thickness values of each 3D slice were documented in a coordinate system chart to generate a scatter graph. Graphs were embedded into images (Orange software) and images were analyzed via hierarchical clustering. RESULTS: Three STS groups were identified according to shape. Two shapes corresponded to the maxilla defects: a square (n = 13) with dimensions of 10 mm in a lingual-buccal (length) and 7-10 mm in a mesial-distal (width) direction; a rectangle (n = 11) of 11 mm in length and 4-7 mm in width. The average shape for mandible defects (n = 9) was smaller (6-8 mm in length, 5-10 mm in width). The highest thickness in all STS was in the buccal portion, above the alveolar ridge, with median values of 2 mm. The lowest thickness of 0.2 mm was at the edges. CONCLUSIONS: The study developed novel methodology to design customized, as well as average shape STS for volume augmentation. Future STS harboring adapted geometry might increase volume augmentation efficiency and accuracy, while reducing surgical time.


Subject(s)
Alveolar Ridge Augmentation , Tooth , Computer-Aided Design , Esthetics, Dental , Humans , Mandible/surgery , Maxilla/surgery
19.
Materials (Basel) ; 14(12)2021 Jun 18.
Article in English | MEDLINE | ID: mdl-34207110

ABSTRACT

The aim of this study was to investigate the effects of saliva contamination and the cleaning of the bond surface of titanium base (ti-base) abutments on the bonding stability and retention force values. The bond surface of the ti-base abutments was treated with airborne-particle abrasion. After contamination, the ti-base abutments underwent different cleaning protocols: water spray (H2O); alcohol (ALC); suspension of zirconium particles (SZP); reapplied airborne-particle abrasion (APA); and a control condition without contamination and cleaning (CTR). All lithium disilicate crowns were bonded to the ti-base abutments using a primer and a self-curing composite. Bonded specimens underwent thermo-mechanical aging. Bond failure analysis and pull-off testing were performed. Bond failure occurred more frequently in groups H2O, ALC, SZP, and APA (p < 0.05). Significant differences in retention force values were only found between CTR and ALC (p < 0.05). Specimens which did not show bond failure after ageing had higher retention force values than the specimens that showed bond failure (p < 0.05). Saliva contamination with cleaning can degrade the bonding properties to titanium. For the retention force values, only the protocol with alcohol after contamination could not restore the values.

20.
Materials (Basel) ; 14(9)2021 Apr 23.
Article in English | MEDLINE | ID: mdl-33922688

ABSTRACT

A novel ceria-stabilized zirconia-alumina-aluminate composite (Ce-TZP-comp) that is not prone to aging presents a potential alternative to yttrium-stabilized zirconia for ceramic oral implants. The objective of this study was to evaluate the long-term stability of a one-piece narrow-diameter implant made of Ce-TZP-comp. Implant prototypes with a narrow (3.4 mm) and regular (4.0 mm) diameter were embedded according to ISO 14801, and subgroups (n = 8) were subsequently exposed to dynamic loading (107 cycles, 98N) and/or hydrothermal treatment (aging, 85 °C). Loading/aging was only applied as a combined protocol for the 4.0 mm diameter implants. One subgroup of each diameter remained untreated. One sample was cross-sectioned from each subgroup and evaluated with a scanning electron microscope for phase-transformation of the lattice. Finally, the remaining samples were loaded to fracture. A multivariate linear regression model was applied for statistical analyses (significance at p < 0.05). All samples withstood the different loading/aging protocols and no transformation propagation was observed. The narrow diameter implants showed the lowest fracture load after combined loading/aging (628 ± 56 N; p < 0.01), whereas all other subgroups exhibited no significantly reduced fracture resistance (between 762 ± 62 and 806 ± 73 N; p > 0.05). Therefore, fracture load values of Ce-TZP-comp implants suggest a reliable intraoral clinical application in the anterior jaw regions.