Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters










Database
Language
Publication year range
1.
Anaesthesia ; 64(11): 1229-35, 2009 Nov.
Article in English | MEDLINE | ID: mdl-19825059

ABSTRACT

The Zeus anaesthesia machine includes an auto-control mode which allows targeting of end-tidal volatile and inspired oxygen concentrations. We assessed the clinical benefits and economic impact of this target-controlled anaesthesia compared with conventional manually controlled anaesthesia. Eighty patients were randomly assigned to receive desflurane either with a fresh gas flow set by the anaesthetist or in auto-control mode. Drug delivery was adjusted to maintain bispectral index between 40-60 units and systolic arterial pressure under 15 mmHg above its pre-induction value (upper limit) and over 90 mmHg (lower limit). Blood pressure was maintained in the desired range for 89% and 91% of the maintenance period for auto-control and manual control respectively (p = 0.49). Bispectral index was in the desired range for 82% and 79% of the maintenance period, for auto-control and manual control respectively (p = 0.46). Oxygen consumption was more than halved by the use of auto-control mode, and mean (SD) desflurane consumption during surgery was 0.07 (0.04) vs 0.2 (0.07) ml.min(-1) in auto-control and manual control respectively (p < 0.0001). The number of drug delivery adjustments per hour was significantly lower in auto-control mode (mean (SD) 7 (2) vs 15 (12); p < 0.0001). Thus, the auto-control mode provided similar haemodynamic stability and bispectral control as did conventional manually controlled anaesthesia, but led to a reduction in gas and vapour consumption with a more clinically acceptable workload.


Subject(s)
Anesthesia, Inhalation/instrumentation , Anesthetics, Inhalation/administration & dosage , Drug Delivery Systems/instrumentation , Isoflurane/analogs & derivatives , Adolescent , Adult , Aged , Anesthesia, Inhalation/methods , Blood Pressure/drug effects , Clinical Protocols , Desflurane , Drug Administration Schedule , Electroencephalography/drug effects , Humans , Isoflurane/administration & dosage , Middle Aged , Monitoring, Intraoperative/methods , Prospective Studies , Workload , Young Adult
2.
Br J Anaesth ; 94(3): 306-17, 2005 Mar.
Article in English | MEDLINE | ID: mdl-15591326

ABSTRACT

BACKGROUND: The aim of this study was to detail the time-course, defined as the changes in end-tidal drug concentration with time, and consumption of inhaled anaesthetics when using a multifunctional closed-circuit anaesthesia machine in various drug delivery modes, and to compare it with a classical anaesthesia machine using an out-of-circle vaporizer under high and low fresh gas flow conditions. METHODS: Using an artificial test lung, sevoflurane and desflurane time-course and consumption were compared when using the Zeus apparatus (Dräger, Lubeck, Germany) with direct injection of inhaled anaesthetics or the Primus apparatus (Dräger, Lubeck, Germany) using a classical out-of-circle vaporizer. Anaesthetics were targeted at 1 and 2 MAC end-tidal during 15 min. For both apparatus, out-of-circle high and low fresh gas control (FGC) and for Zeus, auto-control (AC) modes (fixed fresh gas flow at 6 and 1 litre min(-1) and uptake mode) were compared. Time to reach target, initial overshoot and stability at target, and wash-out times were compared. RESULTS: In FGC, an initial overshoot in end-tidal drug concentration is seen when using 6 litre min(-1) fresh gas flow and a slower time course is observed when using only 1 litre min(-1) in both apparatus. In auto-control mode, the time course of both sevoflurane and desflurane was very fast and not influenced by the changes in fresh gas flow. No overshoot at target was seen. At all settings, the wash-out times were faster when using Zeus than Primus. Inhaled anaesthetic consumption was lowest with the Zeus ventilator in uptake AC mode. CONCLUSION: A combination of the fastest time course and lowest consumption of sevoflurane and desflurane was found when using the Zeus apparatus in AC uptake mode.


Subject(s)
Anesthesia, Closed-Circuit/instrumentation , Anesthetics, Inhalation/administration & dosage , Drug Delivery Systems/instrumentation , Isoflurane/analogs & derivatives , Ventilators, Mechanical , Desflurane , Drug Administration Schedule , Drug Delivery Systems/methods , Electronics, Medical , Equipment Design , Humans , Isoflurane/administration & dosage , Methyl Ethers/administration & dosage , Sevoflurane