Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Langmuir ; 28(33): 12151-8, 2012 Aug 21.
Article in English | MEDLINE | ID: mdl-22831705

ABSTRACT

A quartz crystal microbalance (QCM) sensor was developed for the quantitation of calcium phosphate mineralization and the assessment of DNA as a template molecule. Inherent advantages of QCM, such as nanogram sensitivity, temporal resolution, surface-based measurements, and flow capabilities, were leveraged in the design of this sensor, and in-line fluidic mixing was used to control precursor reaction. This research shows that DNA, a highly programmable anionic polymer, is able to template and control mineralization of calcium phosphate, with nucleation occurring in less than 15 min and initial rates ranging from 4 to 8 ng/min. FT-IR measurements show mineralized material to be calcium phosphate resembling hydroxyapatite (HAP) when a DNA template is used. DNA is a promising mineralization template, and the QCM proves to be a dynamic technique for a broad range of heterogeneous mineralization experiments in complement to classic, diffusion-limited, end-point analysis techniques.


Subject(s)
Calcium Phosphates/metabolism , DNA/metabolism , Minerals/metabolism , Quartz Crystal Microbalance Techniques , Base Sequence , Biosensing Techniques , Calcium Chloride/metabolism , DNA/genetics , Phosphates/metabolism
SELECTION OF CITATIONS
SEARCH DETAIL
...