Biomed Phys Eng Express ; 8(2)2022 01 28.
Article in English | MEDLINE | ID: mdl-35039465

ABSTRACT

By using the statistical techniques of the ANOVA means test and regression, it was found that theNKRcalibration factor of Standard Imaging (SI) model HDR 1000 plus chambers presents a quadratic dependence with the Reference air kerma rateKR(from 6.9 mGy h-1to 43.9 mGy h-1). In order to understand and correct this dependency one model is presented for total recombination:ks=I300I150=1+kini+kd+kvol·I300+kscreen·I3002,wherekiniis the initial recombination,kvolthe thermal diffusion recombination,kvolthe volumetric recombination andkscreenthe screening for the currents/charges collected at the potential differences of 300 and 150 V. In conclusion, the total recombinationksis composed by onekiniwith a constant contribution of 0.019%, onekdcontribution of 0.017%, onekvol·I300contribution from 0.022% to 0.138%, and thekscreen·I3002effects from 0.002% to 0.09% in the range ofK̇Rrate above. However, when this model forksis applied to try to correct the quadratic dependence of theNKRversusKR,explicitly there is no improvement in the variation range of 0.5% of theNKRversusKR.Nonetheless, it allows to obtainNKRvalues consistent with auc ≤ 0.7%, which is less than 1.25% reported in the literature by ADCLs or SSDLs.